| 1301 | Accidentally left crystal biased for a long time. Flash for 1 minute. Before flash, zero Q bias and temp. rises to ~41 mK, then settles back down. Weird. No discernible reason… Flash! |
| 1350 | Leak-checked LHe transfer tube vacuum jacket. The leak seen previously at the valve was not present. The leak at the joint between the fridge-side tube and the flexible section returned (maybe the vacuum grease finally got sucked through). Applied some Q-compound to this joint. Before the Q-compound the steady-state leak was 20 pbarL/s, after applying it the rate fell to 1 pbarL/s. |
| 1400 | Boards in bad state - No pulses and tower temp. ~5 mK warmer. Unplug then replug boards - Temp. decreases. |
| 1418 | Commence LHe transfer (9.8%). |
| 1504 | Finished LHe transfer (91.3%). |
| 1510 | ZAP all SQUIDS and reapply settings; Change C2 driver offset from -0.014 V to 0.038 V; Increase C1 threshold from 25 to 35 ADC bias; Change C1 driver offset from -0.006 V to -0.036 V. |
| 1527 | Take 30 min. of -4/+4 V data (background). |
| 1605 | Stop data run, zero Q bias, flash for 10 sec. |
| 1612 | Flashed for I'm not sure how long… 60 - 70 seconds, I think. Took about 16 min. to cool. |
| 1702 | Stop data run, zero Q bias, flash 10 seconds. Cool down takes 16 minutes. |
| 1706 | Fill LN2 cold trap full. |
| 1723 | Start data run. |
| 1750 | Stop data run. Zero Q. Place Cf source in shield at detector level. |