| 1520 | Putting the mix back into the line. Leak rate increased on e-7 level. from 5.4e-6 to 5.6 e-6. |
| 1530 | valve out LD. See if temperatures change: 1K = 10.9 K, MC(Cernox) = 20.9 K, still = 24.9 K. |
| 1630 | temperature is slowly dropping: 1K = 10.55 K, MC(Cernox) = 20.79 K, still = 23.08 K. Could be either because of lack-of-turbo cooling, or mixture leak. |
| 1630 | open to LD again. Jump to 6e-5 first, then slowly die down to 5e-5, and slowly to 3e-5. |
| 1635 | collecting mix. |
| 1640 | 1K = 11.83 K (why the jump??), MC = 20.68 K, still = 22.84 K |
| 1645 | next I will keep the mix in the dump, and valve out the LD for an hour, and see what happens. |
| 1745 | leak rate back to 5.7e-6. |
| 1745 | mix back into the dump. Close all GHS valves. stop 3He pump. |
| 1745 | temperatures: 1K = 14.55 K, MC(Cernox) = 20.37 K, still = 22.98 K. still temperature stopped decreasing. |
| 1745 | Now valve out LD. |
| 1845 | 1K = 14.78 K, MC(Cernox) = 20.3 K, still = 24.67 K. |
| 1845 | open IVC to LD, leak rate: 5e-5, then decreasing slowly to 3e-5. Very similar behaviour compared to when mix is in the line. |
| 1850 | slightly open 1K needle valve. No change of leak rate. |
| 1855 | introduce He exchange gas. temperatures started dropping. |
| 1925 | MC(Cernox) is going down at a rate of 0.9 K / 2 min. Still started with 1K/2 min but is plateaued at 15 K. why??? 1K is also plateauing at 8.6 K |
| 1925 | MC(Cernox) = 11.3 K, 1K = 8.7 K, still = 15.6 K. |
| 1932 | MC(Cernox) = 9.8 K, start pumping on IVC. |