2025/07/24 (Thursday)

Previous Day

  • Lab in-personnel: Himangshu, Yan
Status 1K valve leak checked and leak rate < 1e-6 mbar L/s. Pumped out IVC and CVJ overnight.
1411 Cryocleaning ongoing. Contaminants from yesterday ~ 150 mbar. Pumped out with scroll pump. Restarted another cycle
1412 Reconnect exchange gas port to turbo pump Tee assembly.
15xx Fixed the main bath He level meter. the power button needs to be pulled out before pressed in. Also, resoldered the V– wire on the DB9 connector.
1622 Collected mixture. Heat gun on the cold trap. reading ~200 mbar. Pumped out with scroll. Doing another cycle
1822 Collected mixture. Heat gun on the cold trap. reading ~100 mbar. Pumped out with scroll.
17xx Flushed main bath and 1k pot with helium to 500 mbar. pumped out to <1 mbar with scroll.
1830 Flushed main bath and 1k pot with helium to 500 mbar again. pumped out to < 1 mbar with scroll.
1840 Put >1 bar of helium in bath and pot for overpressure overnight. Closed main bath valve, speedivalve, and needlevalve.
1855 Doing another cycle of cryocleaning.
1900 Also did electrical continuity check for all three cables (see results below). Only did squid/feedback/qet check. Did not do LED, heater, source mover check.
2000 Pump mixture back into the dump. Closed V3, V5, and V2. Turn off circulation pump. Took out the cold trap and let it warm to room temperature.
17xx Starting electric check.
Detector: S104 Stripline 48
SC: none 300 K 300 K
Face: 3 Date 07/24/2025
SQUET: ZC-39 Cable 1
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 4.2 k OK
SQUID bias B/Ret 19 34 3.96 k OK
SQUID bias C/Ret 8 24 — k 20 nS
SQUID bias D/Ret 23 39 — k OK
Feedback A/Ret 37 21 15 - 25 k 24.58 k 65 nS
Feedback B/Ret 2 18 23.21 k OK
Feedback C/Ret 41 25 25.40 k short
Feedback D/Ret 22 6 16.39 k 35 nS
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 119, 65, 61, 63, 63 ?
QET B/Ret 35 29, 33, 47, 48 78, 77, 119, 79 ?
QET C/Ret 7 5, 9, 10, 28 106, 67, 57, 104 ?
QET D/Ret 38 36, 40, 43, 44 75, 75, 75, 75 ?
QET A/QET B 3 35 1000-3000 1984
QET A/QET C 3 7 2662
QET A/QET D 3 38 2393
QET B/QET C 35 7 2767
QET B/QET D 35 38 2482
QET C/QET D 7 38 1496
1557 Electrical check of ZC-70+interposer
Cooldown checkout sheet (rev C SQUET)
Detector: S104 Stripline 41
SC: none 300 K 300 K
Face: 5 Date 07/24/2025
SQUET: ZC-70 Cable 2
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k — k OK
SQUID bias B/Ret 19 34 4.3 k 5 Mohm, Squid and Feedback is shorted by 6 kOhm
SQUID bias C/Ret 8 24 4.5 k 4 Mohm, Squid and Feedback is shorted by 9 Mohm
SQUID bias D/Ret 23 39 5.0 k OK
Feedback A/Ret 37 21 15 - 25 k – k OK
Feedback B/Ret 2 18 25.0 k 5 Mohm
Feedback C/Ret 41 25 28 k 6 Mohm
Feedback D/Ret 22 6 34 k OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 OL, 95, OL, OL, OL 4 Mohm
QET B/Ret 35 29, 33, 47, 48 108, 100, OL, OL 4 Mohm
QET C/Ret 7 5, 9, 10, 28 109, 79, OL, OL 4 Mohm
QET D/Ret 38 36, 40, 43, 44 79, 80, OL, 4 Mohm 4 Mohm
QET A/QET B 3 35 1000-3000 17.9 kohm ???
QET A/QET C 3 7 800
QET A/QET D 3 38 1947
QET B/QET C 35 7 17.4 kOhm ???
QET B/QET D 35 38 18.8 kOhm ???
QET C/QET D 7 38 1691
LED 1 chassis 42 1.0 - 1.1 V 1.105 1.105
LED 2 chassis 26 1.094 1.69
Motor Coil Y1 pBr3 pBr4 0.249 k
Motor Coil Y2 pCr3 pCr4 0.194 k
Motor Coil X1 pAr3 pAr4 0.186 k
Motor Coil X2 pAr5 pDr3 0.171 k
Xendstop pAr1 chassis open
Yendstop pDr4 chassis 72 ohm
1931 3rd Electrical check of ZC-13 after connecting to Stripline 43 on Tower Face 2
Cooldown checkout sheet (rev C SQUET)
Detector: S104 Stripline 43
SC: none 300 K 300 K
Face: 2 Date 07/24/2025
SQUET: ZC-13 Cable 3
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 4.584 k 39 nS
SQUID bias B/Ret 19 34 4.758 k OK
SQUID bias C/Ret 8 24 — k OK
SQUID bias D/Ret 23 39 4.881 k OK
Feedback A/Ret 37 21 15 - 25 k 18.78 k OK
Feedback B/Ret 2 18 19.84 k OK
Feedback C/Ret 41 25 — k OK
Feedback D/Ret 22 6 20.58 k OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 128.7, 76.5, 72.6, 72.9, 86.3 4 Mohm
QET B/Ret 35 29, 33, 47, 48 89.6, 90.0, 149.1, 89.8 4 Mohm
QET C/Ret 7 5, 9, 10, 28 94.2, 74.6, 63.8, 72.1 4 Mohm
QET D/Ret 38 36, 40, 43, 44 86.4, 75.3, 74.2, 74.6 4 Mohm
QET A/QET B 3 35 1000-3000 1652
QET A/QET C 3 7 2005
QET A/QET D 3 38 1712
QET B/QET C 35 7 1860
QET B/QET D 35 38 1791
QET C/QET D 7 38 2142