| 5:20 | Pressure on G2 is 0.1 mbar during leak checking |
| 5:30 | We think the baseline leak rate is high because the main bath has been soaked in helium for a long time (>10 years), which created a high He background. We decided to try out luck and spray He with the lowest baseline leak rate configuration. |
| 5:30 | With main bath closed (pumping on only 1K pot), needlevalve closed, we spray He on top of the needle valve. |
| 6:40 | Wrapped plastic around pressure gauge housing, attached leak detector to turbo pump Tee; Leak rate background = 4e-8 mbarL/s. With Helium sqirting inside the plastic bag, LD 5.1e-8 L/s. It went upwards very slowly, not consistent with the type of leak we are expecting. The pressure also didn't jump when Turbo is not pumping the volume. Looks like we see it only when we blanked it off. |