| 1225 | Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 100 mbar on G3. |
| | Pump out 1K Pot slowly, throttling the flow at the valve at the pump inlet, opening just until the pump gets louder, then waiting, then opening a little more, etc. With pump just a little loud, can see pressure go down at the mechanical gauge. When this is low, open valve a bit more. When P2 starts to lower down, open valve all the way. |
| 1239 | Open 1K Pot needle valve → 15°. 1K Pot → – K / 20 mbar, Still = 11.1 K, Coldplate = 11.07 K |
| 1248 | Still = 7.94 K, Colplate= 10.4 K, MC temp dropping linearly. |
| 1254 | Still = 5.05 K, Colplate= 9.96 K, MC temp dropping linearly. MC = 6.4 K |
| 1258 | 1K Pot needle valve → 10°. Trying to keep P1 closer to 10 mbar as we used to do in Run 76. P2 = 15 mbar |
| 1315 | Open 1K Pot needle valve → 15°. Still warmed up a bit Still=7.07 K, MC flat. |
| 1323 | Still = 4.38 K, Colplate= 9.19 K, MC temp dropping linearly. MC = 6.03 K |
| 1333 | Still = 3.49 K, Colplate= –K, MC temp dropping linearly. MC = 5.9 K |
| 1333 | Open 1K Pot needle valve → 30°(1 unit). P2 = 40 mbar |
| 1345 | Still = 3.26 K, Colplate= –K, MC temp dropping linearly. MC = 5.59 K |
| 1442 | Still = 2.96 K, Colplate= 3.99 K, MC temp dropping almost flat. MC = 2.94 K |
| 1450 | Close V5 and V9, leaving 150 mbar mixture in dump (G2) |
| 1456 | Turn on 3He pump. G1(condenser) → 250 mbar |
| 1501 | Start circulation through Trap 1. VT →openeing very slowly. Condenser peaks at 450 mbar. MC(H35) → 2.01 K. Still → 2.05 K / – mbar |
| 1516 | VT → more slowly. Condenser peaks at 500 mbar. MC(H35) → 1.33 K. Still → 1.84 K / 40 mbar |
| 1524 | After a few turns, G1 = 450 mbar, p1 = 1 atm, MC(H35) = 1.24 K, Still = 2.04 K |