2024/02/08 (Thursday)

Previous Day

  • Lab in-personnel: Himangshu
1200 Refilled Cold Trap 1 from LN2 dewar
Found that DCRC5 was turned “yes” which was causing the DAQ issue yesterday
1340 Cable 1, DCRC4(SQuid Bias: -91.7, -80.8, 0, 0 LockPoint: -9.45, -3.12, -4.21, -7.36 ) and Cable 3, DCRC3(SQuid Bias: -70.4, -100.5, 0, -75.4 LockPoint: -4.55, -13.15, 0, -2.64 )
1358 Realized DCRC3 cannot trigger on channels. Sp, replace it with DCRC1
Cable 1, DCRC4(SQuid Bias: -91.7, -80.8, 0, 0 LockPoint: -9.45, -3.12, -4.21, -7.36 ) and Cable 3, DCRC1(SQuid Bias: -70.4, -100.5, 0, -75.4 LockPoint: -4.55, -13.15, 0, -2.64 )
1500 Pulses were lagging a lot, so connected DCRC3 in loop but not used for any cables
1543 Trigger Cable3 pB == Side2pA on DCRC1pB, Trigger Threshold on DCRC1pB=35
1907 Stopping data taking. About ~3 hrs of 0V data, source at center, trigger on outer channelA
Channel Mapping
Cable 1 → Zip4, Cable 3 → Zip1
Cable 1 → pA, pB, pC, pD → Side2pC, Side2pF, Side1pC, Side1pD → Zip4
Cable 3 → pA , pB, pC, pD → Side2pB, Side2pA, Side1pB, Side2pD → Zip1