| 1430 | Start removing screws to lower main bath. Loosen medium screws at bottom, two large screws at top, then remove all small screws at top. The shortened 5/16 Allen key that's necessary to remove two screws that are located under fridge insert brackets is kept in the plastic box with the Dewar screws. |
| 1505 | Completely removed main bath screws and lowered main bath |
| 1514 | Check contents of warm LN2 Trap 1: 97 (99.7 on G1) mbar. Pump it out with the scroll pump. |
| 1522 | G1 = 2.8 mbar and stable, stop pumping on trap 1. |
| 1638 | Start slowly letting N2 gas into IVC. Hook up the N2 line to the gauge and it reads about 10 mminHg, slowly open the valve so that the gauge slowly creeps up(currently at 5.4e1 mbar on the digital reading), wait until it the pirange gauge goes upto -1 mminHg. |
| 1652 | IVC is up to 1atm of N2. MC = 283.3 K. |