| | Y coils: 30.6 ohm, 21.1 ohm , X coils: 31.1, 31.6 ohm |
| | Setting Absolute position of source doesn't lead to change any temperature. |
| 1708 | Setting X0 Y96 → X-5 Y96 leads to Tower temp at 4.2K and cooling after wards. Xendstop = 14.1 ohm, Yendstop= 15.5 ohm |
| 1712 | Tower below 1K, make another step |
| 1713 | Set X-10 Y96 , temp jumped to 3.8K and cooling,Xendstop = 14.1 ohm, Yendstop= 15.5 ohm |
| 1723 | Tower below 1K, make another step.X-18 Y96, Xendstop = 14.1 ohm, Yendstop= 15.5 ohm. Tower temp at 4.5K |
| 1733 | Tower below 1K, make another step.X-23 Y96, Xendstop = 14.1 ohm, Yendstop= 15.5 ohm.Tower temp at 2.7K |
| 1740 | Tower below 1K, make another step.X-27 Y96, Xendstop = 14.1 ohm, Yendstop= 15.5 ohm.Tower temp at 2.8K |
| 1746 | Tower below 1K, make another step.X-32 Y96, Xendstop = 14.1 ohm, Yendstop= 15.5 ohm.Tower temp at 3.3K |
| 1752 | Tower below 1K, make another step.X-42 Y96, Xendstop = 14.1 ohm, Yendstop= 15.5 ohm.Tower temp at 3.4K |
| 1758 | If I assume that for some reason the X arm is already touching the endstop, then an opposite movement will lead to change in Xendstop value. Tower below 1K, make another step.X-27 Y96, Xendstop = 14.1 ohm, Yendstop= 15.5 ohm.Tower temp at 3.4K |
| | the conclusion will be that the endstop ground loop is not letting me check the resistances. While opening up, check whether the X arm is at the endstop or not. |