2023/11/09 (Thursday)

Previous Day

  • Lab in-personnel: Himangshu, Ping
Status Run 78 in progress.
Swapped DCRC3 to Cable2+diverter board and DCRC1 to Cable 1. This step was done by Matt in Run 76 to make the channels in Cable 1 show response to QET bias change
1441 Tower(U28)/MC(H35)/Coldplate = 96.2/38.1/185 mK, Still(P1) = 1.34 K / 0.13 mbar / 0 mW, G1/G2 = 184/176 mbar, 1K Pot(P2) = 37.5° / 1.63 K / 22 mbar, MB = 23.9%
1445 Fill LN2 cold trap 1 from dewar
1458 Begin LHe transfer @ MB = 23.3%
1506 Lowered the transfer stick further down into fridge. Raised the telescoping section and lowered the transfer stick all the way down into the He dewar.
1512 Turned on the heater to 4 Psi
1545 End LHe transfer @ MB = 94.7%