2023/10/20 (Friday)

Previous Day

* Lab in-personnel: Zach

1431 Check contents of warm LN2 Trap 1: 15.1 on G1. Pump it out with the scroll pump.
1435 G1 = 2.3 mbar and stable, stop pumping on Trap 1
1436 Insert cold trap 1 into and begin filling the LN2 cold trap dewar with LN2
1437 Begin cryocleaning mixture through cold trap 1 following the procedure in procedure:cryoclean_mixture
1441 Open V6 to 11.0% to start mixture flowing
1446 Open V6 to 11.9% to restart flow
1452 Open V6 to 12.1% to restart flow
1544 Open V6 to 12.2% to restart flow
1559 End cryocleaning mixture through trap 1. Pump mixture to dump
1612 P1 = 0.085 mbar, G1 = 1.9 mbar and stable. Close all valves and turn off 3He pump, leaving 716.1 mbar in dump
1616 Remove cold trap 1 from the LN2 bath to warm up overnight