2023/05/31 (Wednesday)
- Lab in-personnel: Himangshu, Ping
| 1444 | Electrical check of ZC-70+interposer after connecting to Stripline 41 on Tower Face 5 | ||||
|---|---|---|---|---|---|
| Cooldown checkout sheet (rev C SQUET) | |||||
| Detector: | S104 | Stripline | 41 | ||
| SC: | none | 300 K 300 K | |||
| Face: | 5 | Date | 04/24/2023 | ||
| SQUET: | ZC-70 | Cable | 2 | ||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis |
| (Gnd lead) | |||||
| Nominal: | |||||
| < 0.1 nS | |||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | OL | OK |
| SQUID bias B/Ret | 19 | 34 | 4.276 k | 2.6 NOT OK | |
| SQUID bias C/Ret | 8 | 24 | 4.546 k | 2.5 NOT OK | |
| SQUID bias D/Ret | 23 | 39 | 5.039 k | OK | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | OL | OK |
| Feedback B/Ret | 2 | 18 | 25.34 k | 2.6 NOT OK | |
| Feedback C/Ret | 41 | 25 | 28.07 k | 2.4 NOT OK | |
| Feedback D/Ret | 22 | 6 | 34.13 k | OK | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | OL, 82.1, OL, OL, OL | 2.5, NOT OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 78.3, 78.6, OL, OL | 2.5, NOT OK | |
| QET C/Ret | 7 | 5, 9, 10, 28 | 76.0, 74.5, OL, OL | 2.5, NOT OK | |
| QET D/Ret | 38 | 36, 40, 43, 44 | 74.4, 75.0, OL, OL | 2.5, NOT OK | |
| QET A/QET B | 3 | 35 | 1000-3000 | 931 ohm | |
| QET A/QET C | 3 | 7 | 754 | ||
| QET A/QET D | 3 | 38 | 2146 | ||
| QET B/QET C | 35 | 7 | 587.6 | ||
| QET B/QET D | 35 | 38 | 2067 | ||
| QET C/QET D | 7 | 38 | 1905 | ||
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.106 | |
| LED 2 | chassis | 26 | 1.8 | ||
| Motor Coil 1 | pBr3 | pBr4 | 7.66 k | 65.8 ohms | |
| Motor Coil 2 | pCr3 | pCr4 | 7.47 k | 37.5 ohms | |
| Near Endstop | pAr3 | chassis | 5.79 k | ||
| Far Endstop | pAr4 | chassis | 22.51 k | ||
| Thermometer ROX | pAr1 | pDr4 | 1.171 k | OK |
| 1453 | Electrical check of ZC-13 after connecting to Stripline 43 on Tower Face 2 | ||||
|---|---|---|---|---|---|
| Cooldown checkout sheet (rev C SQUET) | |||||
| Detector: | S104 | Stripline | 43 | ||
| SC: | none | 300 K 300 K | |||
| Face: | 2 | Date | 4/24/2023 | ||
| SQUET: | ZC-13 | Cable | 3 | ||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis |
| (Gnd lead) | |||||
| Nominal: | |||||
| < 0.1 nS | |||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.587 k | 2.9, NOT OK |
| SQUID bias B/Ret | 19 | 34 | 4.736 k | 0.3, OK | |
| SQUID bias C/Ret | 8 | 24 | – k | 2.1, NOT OK | |
| SQUID bias D/Ret | 23 | 39 | 4.870 k | 0.3 OK | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 18.79 k | 0.3, OK |
| Feedback B/Ret | 2 | 18 | 19.85 k | 0.2, OK | |
| Feedback C/Ret | 41 | 25 | — k | 0.2, OK | |
| Feedback D/Ret | 22 | 6 | 20.60 k | 0.3, OK | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 93.3, 71.0, 68.0, 68.4, 79.9 | 2.8, NOT OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 74.6, 74.8, 119.3, 74.7 | 2.7, NOT OK | |
| QET C/Ret | 7 | 5, 9, 10, 28 | 71.4, 69.6, 60.5, 69.0 | 2.7, NOT OK | |
| QET D/Ret | 38 | 36, 40, 43, 44 | 88.1, 71.2, 70.3, 70.6 | 2.6, NOT OK | |
| QET A/QET B | 3 | 35 | 1000-3000 | 782 | |
| QET A/QET C | 3 | 7 | 1662 | ||
| QET A/QET D | 3 | 38 | 1343 | ||
| QET B/QET C | 35 | 7 | 1790 | ||
| QET B/QET D | 35 | 38 | 1615 | ||
| QET C/QET D | 7 | 38 | 2057 | ||
| 1512 | Electrical check of ZC-39 after connecting to Stripline 48 on Tower Face 3 | ||||
|---|---|---|---|---|---|
| Note: This SQUET need to slanted to make QET B working, seems to be affected by the angle between stripline head connector with the SQUET | |||||
| Detector: | S104 | Stripline | 48 | ||
| SC: | none | 300 K 300 K | |||
| Face: | 3 | Date | 4/21/2023 | ||
| SQUET: | ZC-39 | Cable | 1 | ||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis |
| (Gnd lead) | |||||
| Nominal: | |||||
| < 0.1 nS | |||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.212 k | 0.4, OK |
| SQUID bias B/Ret | 19 | 34 | 3.965 k | 0.4, OK | |
| SQUID bias C/Ret | 8 | 24 | OL | 2.7, NOT OK | |
| SQUID bias D/Ret | 23 | 39 | OL k | 2.5 NOT OK | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 24.58 k | 2.7, NOT OK |
| Feedback B/Ret | 2 | 18 | 23.23 k | 0.3, OK | |
| Feedback C/Ret | 41 | 25 | 25.39 k | 2.6, NOT OK | |
| Feedback D/Ret | 22 | 6 | 16.40 k | 2.7, OK | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 102.3, 65.5, 61.4, 61.6, 66.3 | 2.7, NOT OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 67.9, 68.1, 82.7, 68.1 | 2.6, NOT OK | |
| QET C/Ret | 7 | 5, 9, 10, 28 | 65.0, 62.4, 55.0,75.1 | 2.6, NOT OK | |
| QET D/Ret | 38 | 36, 40, 43, 44 | 59.5, 59.9, 58.9, 59.2 | 2.7, NOT OK | |
| QET A/QET B | 3 | 35 | 1000-3000 | 1756 | |
| QET A/QET C | 3 | 7 | 2543 | ||
| QET A/QET D | 3 | 38 | 2282 | ||
| QET B/QET C | 35 | 7 | 2621 | ||
| QET B/QET D | 35 | 38 | 2347 | ||
| QET C/QET D | 7 | 38 | 1418 | ||
| 1614 | Scroll on IVC = 0.14 mbar, switch to turbo once it goes below 0.1 mbar |
|---|