| 1230 | Started refill LN2 on Cold Trap 1 |
| | MB = 33%. So, we will do the warmup tomorrow morning. For now, I see that the shorts are gone on Cable 2 and 3(weird). So, I will take a few more IbIs measurments. |
| 1610 | |
| | Decided Configuration: Cable 2/3/1 → diverter/nothing/nothing → DCRC50(3)/DCRC55(1)/DCRCX(4) |
| | Note that DCRC4 didnt respond to change in QET bias, but when I plugged the same cable to DCRC1 and change QET bias on that particular channel, it was responding fine which means an issue with the DCRC itself |
| | Keeping DCRC1 on Cable 1 since it has the Side2pF channel to check |