| Status | Mixture in fridge, turbo pumping on IVC, Fridge at base temperature |
| | Cable 2/1/3 → diverter/nothing/nothing → DCRC50(3)/DCRC55(1)/DCRCX(4) |
| 1454 | Tower(RoX)/Tower(U28)/MC(H35)/Coldplate = –/122/39/183 mK, Still(P1) = 1.48 K / 0.14 mbar / 0 mW, G1/G2 = 92/91 mbar, 1K Pot(P2) = 45° / 1.60 K / 15 mbar, MB = 22% |
| 1517 | Tried DCRC4 but it is still not responsing. So, I am connecting Cable 3 (instead of Cable 1) to DCRC1 to see the phonon channels in the SQUET. |
| 1524 | pA and pD showing traces but not pB. Zapping the DCRC to see if that helps. |
| 1546 | Unplugged the Cable and checked at the BOB, all three channels, SQUIDS and Feedback coil showing similar resistances. Plug it back to DCRC1 and all three are showing noise traces now |
| 1548 | Refilled LN2 at the cold trap 1 |
| 1541 | LHe transfer started |
| 1620 | After swapping the two DCRCS and Cables, all phonon channels are showing traces, except the non-functional ones. So, we will keep this configuration for this run |
| | Decided Configuration: Cable 2/3/1 → diverter/nothing/nothing → DCRC50(3)/DCRC55(1)/DCRCX(4) |
| 1630 | Setup for 2-sided readout. Will do IbIs in this mode. |
| 1641 | Zap all SQUIDS |