2023/03/13 (Monday)

Previous Day

* Lab in-personnel: Himangshu

1511 Checking SQUETS from Run 76: SQUET 70, 39 and 13. Issues during the Run: 4 phonon channels cannot be unrailed in Side 1. At 13K electrical check, the following was found which explains the problem.
Cable2 SQUID_A reads open. Cable 2 → Stripline 41 → ZC 70
Cable1 SQUID_C and SQUID_D read open. Cable 1 → Stripline 48 → ZC 39
Cable3 SQUID_C and Feedback_C read open. Cable 3 → Stripline 43 → ZC 13
1610 The above connections are still open and donot seem to response to little twist to the pins which indicate loose connection.
1640 Start pumping out IVC with Scroll pump. Switch over to pumping the CVJ. Then, start pumping on both. After the pressure is below 1e-1 mbar, use the turbo to pump further down.
1700 Close helium recovery valve to main bath, open stopper on main bath to check pressure, there is none.