UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20230222

2023/02/22 (Wednesday)

Previous Day

* Lab in-personnel: Himangshu

Status Run 77 in progress.
Cooling to base today
1227 Turbo on IVC = 1.2 nbar, with turbo valved out for a min IVC = 12 nbar
1230 Open up turbo again, close IVC valves on top of fridge
1234 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed).Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.24 bar (-22.8 inHg) on mechanical gauge.
1242 Pump out 1K Pot slowly, opening the 4He pump valve until it just starts to make noise, then opening it more when it becomes more quiet.
1244 Start pumping 1K Pot continuously
1249 Fill LN2 cold trap dewar
1250 MC(H35) = 5.32 K, cooling slowly. 1K Pot = 3.37 K / 1.7 mbar, cooling slowly
1252 Open 1K Pot needle valve → 15° (half a unit mark). 1K Pot (P2) → 3.06 K / 6.0 mbar
1255 After about 3 minutes, MC temperature started dropping MC(H35) = 5.23 K
1256 Stopped filling LN2 dewar
1259 Open 1K Pot needle valve → 30°. 1K Pot → 2.6 K / 18 mbar. MC(H35) = 3.4 K
1304 Open 1K Pot needle valve → 45°. 1K Pot → 1.91 K / 30 mbar MC(H35) = 3.25 K
1306 G1/G2 = 345/334 mbar dropping linearly
1310 MC(H35)/1Kpot = 3.08/1.72K dropping linearly
1327 MC(H35)/1Kpot = 2.38/1.63K still dropping linearly. G1/G2 = 97/116 mbar . 1K Pot(P2) reads 20 mbar.
1345 Repaired the broken wire on the Tower Rox thermometer plug, should take measurement for this thermometer this time
1347 MC(H35)/1Kpot = 2.09/1.557 K still dropping linearly. G1/G2 = 58/61.6 mbar . 1K Pot(P2) reads 8.0 mbar.
1351 Close V3 and V9, leaving 58.6 mbar mixture in dump. P2 jumped up to 32 mbar and coming down slowly.
I will do the rest of the cooldown remotely
1521 Turn on 3He pump
1525 Start circulation through Trap 1. V6 → 12.6%. P2 went to 40 mbar and came back down. Condenser peaks at 298.2 mbar. MC(H35) → 1.39 K. Still(P1) → 1.43 K / 14 mbar. Took about 2 minutes for P1 to start reading below max.
1531 V6 → 13.0%. Condenser(G1) small bump at 252 mbar. MC(H35) → 1.19 K. Still(P1) → 1.21 K / 5.0 mbar.
1537 V6 → 13.4%. Condenser(G1) peaks at 231.3 mbar. MC(H35) → 1.14 K. Still(P1) → 1.18 K / 3.2 mbar.
1544 V6 → 13.8%. Condenser(G1) peaks at 230.8 mbar. MC(H35) → 1.08 K. Still(P1) → 1.23 K / 3.2 mbar.
1557 Power cycling DCRC1 and DCRC4
1602 V6 → 14.4%. Condenser(G1) peaks at 247.5 mbar. MC(H35) → 0.66 K. Still(P1) → 1.19 K / 1.3 mbar.
1612 V6 → 15.0%. Condenser(G1) peaks at 242.9 mbar. MC(H35) → 0.377 K. Still(P1) → 1.18 K / 0.7 mbar.
1621 V6 → 100%. Condenser(G1) peaks at 234.2 mbar. MC(H35) → 0.166 K. Still(P1) → 1.11 K / 0.14 mbar. P2 remain steady at 32 mbar during the process of opening up V6
We were power cycling the DCRCs to see channels live during the above steps
1701 MC(H35) = 33.8 mK and pretty flat, G1 = 163.5 mbar and pretty flat. 1K Pot = 1.66 K / 30 mbar.
1702 Open V14 to pull the rest of the mixture out of the dump into circulation. P1 peaks at 340 and comes back down slowly
1706 MC rises to 88 mK, Still peaks at 263 mbar (on readout) then falls again, Condenser(G1) rises to 253 mbar. 1K Pot unaffected.
1708 MC seems to have a second wind and is rising again to 97.3 mK, condenser is falling like usual
1734 P1 = 0.55 mbar (<1), close V14
From IbIs with MC = 31.8 mK
SQUET Cable DCRC channel Detector channel SQUID Bias SQUID Lockpoint QET bias @ dP/dI = - uV 3rd Pumpout Previous Run
ZC-74 1 DCRC1pC Side1pC -92.6 -3.98 no transition(feedback resistance low) 235
ZC-74 1 DCRC1pD Side1pF -97.9 -2.4 148 274
ZC-31 2 DCRC3pA Side1pE -98.1 -6.12 300 135 246
ZC-31 2 DCRC3pB Side1pB -87.6 -3.49 308 190 225
ZC-31 2 DCRC3pC Side1pD -44.4 -2.86 no transition 258
ZC-26 3 DCRC4pC Side1pA -83.7 -7 320 212
ZC-74 1 DCRC1pA Side2pC -95(copy) -2.5(copy) no transition
ZC-74 1 DCRC1pB Side2pF -100 -2.01 no transition
ZC-31 2 DCRC3pD Side2pE -89.1 -3.81 no transition
ZC-26 3 DCRC4pA Side2pB -83(copy) -5.91(copy) no response to change in QETbias
ZC-26 3 DCRC4pB Side2pA -83(copy) -5.91(copy) no response to change in QETbias
ZC-26 3 DCRC4pD Side2pD -93.1 1.56 no transition

Next Day