2023/02/20 (Monday)

Previous Day

* Lab in-personnel:

Begin another cooldown
Status Run 77 in progress.
Setup To Be Cable 2/1/3 → diverter/nothing/nothing → DCRC50(3)/DCRC55(1)/DCRCX(4)
1636 Started Ln2 Cold trap refill
1638 Tower(RoX)/Tower(U28)/MC(H35)/Coldplate = -/47.4/31.1/152 mK, Still(P1) = 1.14 K / 0.08 mbar / 0 mW, G1/G2 = 145/153 mbar, 1K Pot(P2) = 30° / 1.624 K / 20 mbar, MB = 55%
1641 Turned off DCRC programs. Unplugged DCRC4. No changed in Tower thermometer reading.
1644 Unplugged DCRC1 and DCRC3. The tower temperature dropped a little bit to 43.5 mK.
1648 Plug in DCRC1. No change in tower temperature. Plug in DCRC3. No noticeable change. Plug back DCRC4.
1705 DCRC4 still not responding. Zapping the SQUIDs on this DCRC.
Tower temperature went upto to 50.5 mK and cooling back down. Settled at 48.6 mK