2023/02/19 (Sunday)

Previous Day

* Lab in-personnel: Zach

Begin another cooldown
1132 Turbo on IVC = 1.2 nbar, with turbo valved out for a min IVC = 14 nbar
1133 Open up turbo again, close IVC valves on top of fridge
1137 Begin LHe transfer @ MB = -1.0%
1140 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (open V4a, close 1K pot speedivalve and slowly open the 4He-pump inlet valve).Close the 4He-pump inlet valve once done. Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.23 bar (-23 inHg) on mechanical gauge.
1145 Pump out 1K Pot slowly, opening the 4He pump valve until it just starts to make noise, then opening it more when it becomes more quiet.
1146 Open 4He pump valve fully.
1147 Fill LN2 cold trap dewar
1150 MC(H35) = 5.50 K, 1K Pot = 3.37 K / 2.4 mbar
1155 MB = 36.6%, thats enough. Open 1K Pot needle valve → 15° (half a unit mark). 1K Pot (P2) → 2.77 K / 9.0 mbar
1200 MC temp dropping
1203 Open 1K Pot needle valve → 30°. 1K Pot → 2.07 K / 24 mbar.
1210 MC(H35) = 3.03 dropping linearly. G1/G2 = 275/272 mbar . 1K Pot(P2) = 1.75 K / 24 mbar.
1229 Ended LHe transfer @ MB = 92.5%
1233 P2 dropped to 12 mbar, MC(H35) = 2.24 K, about the transition point
1243 Plug in DCRCs
I will do the rest of the cooldown remotely
1306 MC(H35)/1Kpot = 1.90/1.61 K and flat. G1/G2 = 51/54 mbar . 1K Pot(P2) = 20 mbar.
1307 Close V3 and V9, leaving 53.6 mbar mixture in dump.
1307 Turn on 3He pump
1308 Start circulation through Trap 1. V6 → 12.6%. Condenser(G1) peaks at 326 mbar. MC(H35) → 1.53 K. Still → 1.65 K / 40 mbar.
1317 V6 → 13.0%. Condenser(G1) peaks at 206 mbar. MC(H35) → 1.21 K. Still(P1) → 1.39 K / 5.5 mbar.
1321 V6 → 13.4%. Condenser(G1) peaks at 255 mbar. MC(H35) → 1.17 K. Still(P1) → 1.35 K / 4.5 mbar.
The MC temp went up a bit (to 1.3K) before going down again
1331 V6 → 13.8%. Condenser(G1) peaks at 204 mbar. MC(H35) → 0.94 K. Still(P1) → 1.27 K / 2.4 mbar.
1340 V6 → 14.4%. Condenser(G1) peaks at 210 mbar. MC(H35) → 0.59 K. Still(P1) → 1.30 K / 1.3 mbar.
1347 V6 → 15.0%. Condenser(G1) peaks at 207 mbar. MC(H35) → 0.38 K. Still(P1) → 1.29 K / 0.75 mbar.
1356 V6 → 100.%. Condenser(G1) peaks at 198 mbar. MC(H35) → 0.16 K. Still(P1) → 1.24 K / 0.15 mbar.
Still temp and G1 are kinda high
1427 MC(H35) = 45.8 mK and pretty flat, G1 = 144 mbar and pretty flat. 1K Pot = 1.62 K / 20 mbar.
1427 Open V14 to pull the rest of the mixture out of the dump into circulation.
1428 MC(H35) rises to 91 mK then falls, Still rises to 90 mbar then falls, Condenser(G1) rises to 220 mbar then falls. 1K Pot unaffected.
1449 P1 = 0.95 mbar (<1), close V14
MC and still are settling at 43 mK and 1.26 K, higher than usual

Next Day