  • Lab in-personnel: Himangshu, Zach
Status Run 77 preparation
Setup IVC open. Turbo valved out. tower installation in progress
1107 Did solder pins on ZC74 to reduce pin connection failures after connecting to stripline. Installed the three SQUETS and heatsinks.
1400 Begin installing bottom cap to IVC, perform an electrical check halfway and see that Squet ZC31 with the interposer board has some electrical issues.
1445 When testing the connection directly on the Squet, wiggling certain pins causes the connection to be unstable (seen in feedback A and squid D). Himangshu has a method of remelting the solder holding the pins in without losing the pin. This reseats the pins and fixes any cracks in the solder.
1228 Electrical check of ZC-31 after connecting to Stripline 41 on Tower Face 5
Detector: S104 Stripline 41
SC: none 300 K 300 K
Face: 5 Date 1/27/2023
SQUET: ZC-31 Cable 2
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 4.236 k OK
SQUID bias B/Ret 19 34 4.048 k OK
SQUID bias C/Ret 8 24 4.513 k OK
SQUID bias D/Ret 23 39 4.133 OK
Feedback A/Ret 37 21 15 - 25 k 26.64 k OK
Feedback B/Ret 2 18 26.13 k OK
Feedback C/Ret 41 25 28.06 k OK
Feedback D/Ret 22 6 26.83 k OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 OL,82,OL, OL, OL OK
QET B/Ret 35 29, 33, 47, 48 78, 77, OL, OL OK
QET C/Ret 7 5, 9, 10, 28 77, 76, OL, OL OK
QET D/Ret 38 36, 40, 43, 44 72, 72, OL, OL OK
QET A/QET B 3 35 1000-3000 574
QET A/QET C 3 7 743
QET A/QET D 3 38 2120
QET B/QET C 35 7 495
QET B/QET D 35 38 1939
QET C/QET D 7 38 1892
LED 1 chassis 42 1.0 - 1.1 V 1.105
LED 2 chassis 26 1.66
1241 Electrical check of ZC-26 after connecting to Stripline 43 on Tower Face 2
Detector: S104 Stripline 43
SC: none 300 K 300 K
Face: 2 Date 1/27/2023
SQUET: ZC-26 Cable 3
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 4.726 k OK
SQUID bias B/Ret 19 34 3.99 k OK
SQUID bias C/Ret 8 24 3.519 k OK
SQUID bias D/Ret 23 39 3.384 k OK
Feedback A/Ret 37 21 15 - 25 k 1.092 k OK
Feedback B/Ret 2 18 16.78 k OK
Feedback C/Ret 41 25 21.23 k OK
Feedback D/Ret 22 6 20.63 k OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 158, 119, 111, 112, 120 OK
QET B/Ret 35 29, 33, 47, 48 83, 75, 86, 78 OK
QET C/Ret 7 5, 9, 10, 28 72, 70, 58, 67 OK
QET D/Ret 38 36, 40, 43, 44 79, 66, 66, 66 OK
QET A/QET B 3 35 1000-3000 826
QET A/QET C 3 7 1660
QET A/QET D 3 38 1383
QET B/QET C 35 7 1745
QET B/QET D 35 38 1611
QET C/QET D 7 38 2011
LED 1 chassis 42 1.0 - 1.1 V -
LED 2 chassis 26 -
1248 Electrical check of ZC-74 after connecting to Stripline 48 on Tower Face 3
Detector: S104 Stripline 48
SC: none 300 K 300 K
Face: 3 Date 1/27/2023
SQUET: ZC-74 Cable 1
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 4.761 k OK
SQUID bias B/Ret 19 34 4.491 k OK
SQUID bias C/Ret 8 24 4.396 k OK
SQUID bias D/Ret 23 39 4.312 k OK
Feedback A/Ret 37 21 15 - 25 k 21.03 k OK
Feedback B/Ret 2 18 19.76 k OK
Feedback C/Ret 41 25 18.49 k OK
Feedback D/Ret 22 6 18.79 k OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 77, 65, 58, 59, 68 OK
QET B/Ret 35 29, 33, 47, 48 71, 71, 72, 72 OK
QET C/Ret 7 5, 9, 10, 28 55, 64, 54, 68 OK
QET D/Ret 38 36, 40, 43, 44 58, 58, 57, 58 OK
QET A/QET B 3 35 1000-3000 1735
QET A/QET C 3 7 2535
QET A/QET D 3 38 2166
QET B/QET C 35 7 2626
QET B/QET D 35 38 2243
QET C/QET D 7 38 1079
LED 1 chassis 42 1.0 - 1.1 V -
LED 2 chassis 26 -
1615 We closed the IVC. Begin gently pumping IVC with Scroll Pump. Initially the turbo pump gauge reads 3.2e2 mbar for few minutes then it started going down.
1621 Begin tightening the bolts on the bottom of the IVC more.
1624 We opened the scroll pump all the way at this point.
1630 The turbo gauge reads 1e-1 mbar so we are switching to turbo pumping now. As we open the turbo valve, we are also slowly closing the scroll pump valve. After the scroll pump valve is completely closed, unplug the scroll pump to stop it.
1714 Attach fill funnel to outside of IVC. Connect the leak checker to the IVC and wait 20 minutes to get it ready. The leak detector read ~5x10e-10 mbar L/s , then we Open the valve on the leak detector while showly closing the turbo valve, the lead detector settles at 1.04e-8 mbar L/s.
Squirt He around bottom IVC joint. No leak detected. Switch back to turbo pumping.