2022/10/31 (Monday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on W side of fridge.
NaI array trigger thresholds: 7.0 - 11.0 MeV
Warmed up yesterday for nineteenth 5K pumpout. Turbopumping IVC overnight.
0930 Turbo on IVC = 1.0 nbar. With Turbo valved out for 1 minute IVC = 10 nbar.
0930 Close IVC valves on top of fridge
1004 Begin LHe transfer @ MB = 0.9%
1058 End LHe transfer @ MB = 74.9%
note: Mike had a problem with the storage Dewar in the liquefier room, and needs the 250L Dewar to help cool it back down. So I did a partial transfer with a partially-full 100L Dewar. We'll need to transfer again tomorrow.
I'm planning to let the fridge sit at >4K for a day, start cooling tomorrow, as an experiment. We'll see whether the He buildup begins any sooner if we leave the IVC closed for a day at a temperature above the lambda point. (I'm trying to decide if what we have is actually a superleak, which would not start until we cooled the mixture to superfluid.)
1250 Fill LN2 cold trap bath