start › cdms › k100 › run_diary › 20221022

2022/10/22 (Saturday)

Previous Day

  • Lab in-personnel: Zach
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source in storage in PAN 434.
NaI scan data running overnight
Warmed up last night for sixteenth 5K pumpout.
1018 Turbo on IVC = 1.0 nbar. With Turbo valved out for 1 minute IVC = 10 nbar.
1019 Close IVC valves on top of fridge
1029 Begin LHe transfer @ MB = 6.2%
1042 MB = 21.5%
1045 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed).Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.32 bar on mechanical gauge.
1046 Pump out 1K Pot in stages, follow Matt's new procedure of throttling the flow at the valve at the pump itself, opening just until the pump got louder, then waiting, then opening a little more, etc.
1050 Start pumping 1K Pot continuously
1054 MB = 63.0%
1055 MC = 5.92 K, cooling slowly. 1K Pot = 3.16 K / 2.6 mbar
1056 Open 1K Pot needle valve → 15°. 1K Pot → 2.81 K / 7.5 mbar
1100 Open 1K Pot needle valve → 30°. 1K Pot → 2.87 K / 7.5 mbar
1105 Open 1K Pot needle valve → 45°. 1K Pot → 2.78 K / 9.5 mbar
1113 G1/G2 = 329/319 mbar dropping linearly
1113 MC/1Kpot = 3.16/2.64K dropping linearly
1121 Ended LHe transfer @ MB = 94.8%
1145 Fill LN2 cold trap bath
1245 G1(fridge) = 45 mbar. MC = 1.84 K. 1K Pot = 1.51 K / 6.5 mbar
1250 Close V3 and V9, leaving 56 mbar mixture in dump
1250 Turn on 3He pump. G1(condenser) → 52 mbar
1250 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 232 mbar. MC(H35) → 1.45 K. Still → 1.49 K / 40 mbar
1255 V6 → 13.0%. Condenser(G1) peaks at 233 mbar. MC(H35) → 1.29 K. Still(P1) → 1.32 K / 9 mbar.
1300 1K Pot = 1.56 K / 12 mbar
1300 V6 → 13.4%. Condenser(G1) peaks at 202 mbar. MC(H35) → 1.14 K. Still(P1) → 1.22 K / 5.0 mbar.
1305 V6 → 13.8%. Condenser(G1) peaks at 177 mbar. MC(H35) → 1.04 K. Still(P1) → 1.04 K / 1.6 mbar.
MC bounced around 1.04-1.19 K (Al transition)
1310 V6 → 14.4%. Condenser(G1) peaks at 153 mbar. MC(H35) → 0.81 K. Still(P1) → 1.01 K / 0.9 mbar.
1315 V6 → 15.0%. Condenser(G1) peaks at 153 mbar. MC(H35) → 0.64 K. Still(P1) → 1.05 K / 0.65 mbar.
1320 V6 → 100.%. Condenser(G1) peaks at 158 mbar. MC(H35) → 0.30 K. Still(P1) → 1.03 K / 0.17 mbar.
1335 1K Pot = 1.56 K / 10.5±0.5 mbar. Good - it was flat at 7.5 mbar for a while but now it looks good.
1345 MC = 40 mK. Still = 1.08 K / 0.11 mbar. G1/G2 = 111/126 mbar. 1K Pot = 1.56 K / 10.5±0.5 mbar.
1345 Open V14 to pull the rest of the mixture out of the dump into circulation. Still peaks at 55 mbar, Condenser peaks at 168 mbar, MC peaks at 83 mK, then all relax again. 1K Pot unaffected.
1355 Plug in DCRCs 1 and 3
1410 Still = 0.75 mbar.
Close V14
1550 MC = 28.2 mK. Still = 1.11 K / 0.110 mbar. G1/G2 = 107/122 mbar. 1K Pot = 1.57 K / 10±1 mbar.
1600 Check phonon channels. All but DCRC3pA are close to -0.8 V baselines.
1610 Adjust phonon offsets to get all baselines near -0.80 V
1610 MC = 29.2 mK, it has risen since messing with the phonon channels
1614 Dial HV up to -109.25V nominal, -107.1 V actual, for 15% prebias (using RootSequencer)
1620 Dial HV down to -94V nominal, -91.7V actual
1623 HV = -91.7 V, Vtest = -21.5 mV, Ileak = -23.6 uA, Vdet = -82.3 V (90%).
Note: I'm enabling DCRC4 for consistency in processing and data rate. But the NaI array is not in a fixed range so that data is not relevant in the current data
Since the PuBe source is not in the lab we'll take some more Run68-style background data
1630 DCRC1pA baseline = -0.796 V
1630 Start -82.3 V bg data series 859 good min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pA threshold 2.
Start series with the Start Button
1905 DCRC1pA baseline = -0.801 V
2145 HV = -91.8 V, Vtest = -21.6 mV, Ileak = -23.7 uA, Vdet = -82.3 V (90%). Seeing some instability in the bias that I haven't seen before, jumping around a bit between -91.7 and -91.9 V
2145 DCRC1pA baseline = -0.809 V. It has been going down, maybe the detector is still cooling.