UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20221011

2022/10/11 (Tuesday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
thirteenth 5K pumpout yesterday, turbopumping IVC overnight.
1100 MC/1KPot/Sorb = 5.22/6.12/7.8 K. G1(fridge+dump) = 541 mbar
1105 Turbo on IVC = 1.0 nbar. With Turbo valved out for 1 minute IVC = 10 nbar.
1105 Close IVC valves on top of fridge
1110 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.25 bar on mechanical gauge.
1110 Pump out 1K Pot in stages, alternating 1K Pot Speedivalve and V4a, opening V4a in brief spurts to keep 4He pump pressure low (below 5 mbar).
1115 Start pumping 1K Pot continuously. 1K Pot → 3.22 K / 3.0 mbar
1125 MC starts falling quickly. Fridge pressure falling.
1140 MC drops quickly from 4.1 to 3.3 K
1335 MC/1KPot = 2.46/2.06 K. G1(fridge+dump) = 131 mbar
1340 1K Pot = 0° / 2.04 K / 3.0 mbar
1340 Open 1K Pot needle valve → 15°. 1K Pot → 1.61 K / 6.0 mbar
1345 Open 1K Pot needle valve → 30°. 1K Pot → 1.56 K / 7.5 mbar
1350 Open 1K Pot needle valve → 45°. 1K Pot → 1.57 K / 10 mbar
1425 MC = 1.90 K, it kinked at 2.06 K (lambda point). G1(fridge) = 51 mbar
1510 1K Pot = 1.59 K / 11.5±0.5 mbar
1510 MC = 1.73 K. G1(fridge+dump) = 36 mbar
1515 Fill LN2 cold trap bath
1515 Close V3 and V9, leaving 46 mbar mixture in dump (G2)
1515 Turn on 3He pump. G1(condenser) → 41 mbar.
1520 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 236 mbar. MC(H35) → 1.44 K. Still → 1.48 K / 40 mbar
1535 V6 → 13.0%. Condenser(G1) peaks at 282 mbar. MC(H35) → 1.37 K. Still(P1) → 1.40 K / 17 mbar.
a bit different from usual, MC was stuck around 1.45 K for a while. Falling now.
1545 V6 → 13.4%. Condenser(G1) peaks at 179 mbar. MC(H35) → 1.08 K. Still(P1) → 1.14 K / 3.4 mbar.
MC kinked between 1.19 and 1.06 K, Al transition
1555 V6 → 13.8%. Condenser(G1) peaks at 170 mbar. MC(H35) → 0.80 K. Still(P1) → 1.11 K / 2.0 mbar.
1600 V6 → 14.6%. Condenser(G1) peaks at 176 mbar. MC(H35) → 0.57 K. Still(P1) → 1.10 K / 1.0 mbar.
hmm, that's supposed to be 14.4%. Well whatever.
1610 V6 → 15.0%. Condenser(G1) peaks at 167 mbar. MC(H35) → 0.24 K. Still(P1) → 1.14 K / 0.60 mbar.
1615 V6 → 100%. Condenser(G1) peaks at 157 mbar. MC(H35) → 0.06 K. Still(P1) → 1.09 K / 0.12 mbar.
1635 MC = 43 mK. Still = 1.09 K / 0.11 mbar. G1/G2 = 126/140 mbar. 1K Pot = 1.59 K / 11±1 mbar.
1635 Open V14 to pull the rest of the mixture out of the dump into circulation. Still peaks at 28 mbar, Condenser peaks at 173 mbar, MC peaks at 69 mK, then all relax again. 1K Pot unaffected.
1655 Still = 0.85 mbar.
Close V14
1700 MC = 31.5 mK. Still = 1.08 K / 0.100 mbar. G1/G2 = 122/134 mbar. 1K Pot = 1.59 K / 11±1 mbar.
1700 Plug in DCRCs 1&3 and start DAQ programs

Next Day