UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20221008

2022/10/08 (Friday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Cooled to base two days ago after twelfth 5K pumpout.
1500 MC base 29.7 mK, most spikes peaking below 33 mK, some above 35 mK
1510 MC/Coldplate = 29.7/151 mK, Still = 1.13 K / 0.110 mbar / 0 mW, G1/G2 = 134/123 mbar, 1K Pot = 45° / 1.58 K / 11±1 mbar, MB = 32.6%
1517 Begin LHe transfer @ MB = 32.6%
1525 Fill LN2 cold trap bath
1559 End LHe transfer @ MB = 93.9%

Next Day