2022/10/06 (Thursday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Warmed fridge two days ago for twelth 5K pumpout. Turbopumping IVC for the last two days.
1025 MC/1KPot/Sorb = 5.36/6.98/8.9 K. G1(fridge) = 552 mbar
1025 Turbo on IVC = 1.2 nbar. With Turbo valved out for 1 minute IVC = 12 nbar.
1030 Close IVC valves on top of fridge
1030 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.23 bar on mechanical gauge.
1030 Pump out 1K Pot in stages, alternating 1K Pot Speedivalve and V4a, opening V4a in brief spurts to keep 4He pump pressure low (below 5 mbar).
1035 Start pumping 1K Pot continuously
1100 MC = 3.44 K. Dropped quickly to this point, then started cooling more slowly.
1345 MC/1KPot = 2.55/2.12 K. G1(fridge) = 143 mbar
1345 1K Pot = 0° / 2.11 K / 2.4 mbar
1345 Open 1K Pot needle valve → 15°. 1K Pot → 1.78 K / 5.5 mbar
1355 Open 1K Pot needle valve → 30°. 1K Pot → 1.56 K / 8.0 mbar
1400 Open 1K Pot needle valve → 45°. 1K Pot → 1.57 K / 10 mbar
1405 MC = 2.14 K. G1(fridge) = 70 mbar
1445 MC = 1.87 K, it kinked at 2.1 K (lambda point). G1(fridge) = 47 mbar
1450 Fill LN2 cold trap bath
1500 MC = 1.78 K. G1(fridge) = 40 mbar
1500 Close V3 and V9, leaving 30 mbar mixture in dump (G2)
1500 1K Pot = 1.59 K / 11±1 mbar
1505 Turn on 3He pump. G1(condenser) → 46 mbar.
1505 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 236 mbar. MC(H35) → 1.55 K. Still → 1.51 K / 50 mbar
1510 V6 → 13.0%. Condenser(G1) peaks at 272 mbar. MC(H35) → 1.25 K. Still(P1) → 5.5 mbar.
1515 V6 → 13.4%. Condenser(G1) peaks at 223 mbar. MC(H35) → 1.06 K. Still(P1) → 3.2 mbar.
MC kinked around 1.18 and 1.06 K, Al transition
1520 V6 → 13.8%. Condenser(G1) peaks at 184 mbar. MC(H35) → 0.80 K. Still(P1) → 1.16 K / 2.0 mbar.
1530 V6 → 14.6%. Condenser(G1) peaks at 178 mbar. MC(H35) → 0.46 K. Still(P1) → 1.13 K / 1.1 mbar.
1535 V6 → 15.0%. Condenser(G1) peaks at 181 mbar. MC(H35) → 0.22 K. Still(P1) → 1.16 K / 0.55 mbar.
1546 Begin LHe transfer @ MB = 38.0%
starting transfer causes a bit of a rise in condenser and 1KPot
1555 V6 → 100%. Condenser(G1) peaks at 159 mbar. MC(H35) → 0.14 K. Still(P1) → 1.05 K / 0.17 mbar.
1600 MC = 77 mK. Still = 1.09 K / 0.14 mbar. G1/G2 = 148/136 mbar. 1K Pot = 1.56 K / 11±1 mbar.
1600 Open V14 to pull the rest of the mixture out of the dump into circulation. Still peaks at 40 mbar, Condenser peaks at 192 mbar, MC peaks at 105 mK, then all relax again. 1K Pot unaffected.
1625 Still = 0.90 mbar.
Close V14
1630 End LHe transfer @ MB = 93.6%
1635 MC = 29.7 mK. Still = 1.09 K / 0.110 mbar. G1/G2 = 112/100 mbar. 1K Pot = 1.56 K / 11±1 mbar.
1640 Plug in DCRCs 1&3 and start DAQ programs
1650 had a bit of trouble starting a run and keeping the DAQ programs from crashing. Wound up powercycling the DCRCs, got it to work.