2022/10/04 (Tuesday)

Previous Day

  • Lab in-personnel: Giulia
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Cooled to base three days ago after eleventh 5K pumpout. Taking -83V low-threshold SiHV-triggered PuBe data overnight.
0240 At about this time dump sizes fall. Probably this is when helium films started hurting the phonon channel performance. That's around dump 2990.
0500 Phonon baselines are highly elevated. Stop series.
0500 Dial HV down to 0V
0505 Close V6 to stop circulation
0510 Turn off 3He pump
0955 Close 1K Pot needle valve
1000 Close 1K Pot speedivalve
1055 Turn on Still and MC heaters to 20 mW each
1130 Open V3; G1 rises, so this released pressure on the Still side.
1155 G1(fridge) = 859 mbar.
Open V9 to release mixture into dump.
1330 MC/1KPot/Sorb = 5.80/5.93/7.4 K. These are flat now.
1330 Close V4a, 4He pump inlet valve. Turn off 4He pump. Close recovery valve to 4He pump outlet. NOTE: remember to open recovery valve when the pump is turned back on.
1335 Unplug DCRCs 1&3 (can't turn off DAQ programs, don't have access to cdms1 currently)
1345 Apply some vacuum grease in the gap between the knurled knobs on the stripline elbow, directly to stripline tube. Remove the washers that were holding it up, then push it down so that the knurled knobs are touching.
1345 Check pressure in IVC: 0.10 ubar.
1345 Star pumping out IVC with turbo pump
1500 MC/1KPot/Sorb = 8.01/>7/9.2 K, rising linearly.
1508 Begin LHe transfer @ MB = 35.1%
1515 Fill LN2 cold trap bath
1535 MC/1Kpot/Sorb = 10.03/5.84/7.5 K
Turn off MC and Still heaters
1553 End LHe transfer @ MB = 94.2%

Next Day