start › cdms › k100 › run_diary › 20221001

2022/10/01 (Saturday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe source behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Eleventh 5K pumpout yesterday. Turbopumping IVC overnight.
1615 MC/1KPot/Sorb = 5.40/6.67/8.5 K. G1(fridge) = 545 mbar
1615 Turbo on IVC = 1.4 nbar. With Turbo valved out for 1 minute IVC = 14 nbar.
1615 Close IVC valves on top of fridge
1620 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed). Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.24 bar on mechanical gauge.
1620 Pump out 1K Pot in stages, alternating 1K Pot Speedivalve and V4a, opening V4a in brief spurts to keep 4He pump pressure low (below 5 mbar).
1625 Start pumping 1K Pot continuously
1630 MC/1KPot/Sorb = 5.39/3.25/4.1 K
1630 1K Pot = 0° / 3.25 K / 3.2 mbar
1630 Open 1K Pot needle valve → 15°. 1K Pot → 3.06 K / 6.0 mbar
MC temperature falling
1635 Fill LN2 cold trap bath
1635 MC/Coldplate/Still/Sorb = 3.43/“4.91”/“3.43”/4.0 K
1640 Open 1K Pot needle valve → 30°. 1K Pot → 2.89 K / 7.5 mbar
1640 MC = 3.36 K. G1(fridge) = 422 mbar
1645 Open 1K Pot needle valve → 45°. 1K Pot → 2.60 K / 11 mbar
1645 MC = 3.28 K. G1(fridge) = 389 mbar
1645 Plug in DCRCs 1&3
1745 MC = 2.24 K
1745 Close V3 and V9, leaving 64 mbar mixture in dump (G2)
1745 Turn on 3He pump. G1(condenser) 64 → 90 mbar.
1750 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 303 mbar. MC(H35) → 1.6 K.
1755 V6 → 13.0%. Condenser(G1) peaks at 330 mbar. MC(H35) → 1.35 K. Still(P1) → 30 mbar.
1800 V6 → 13.4%. Condenser(G1) peaks at 285 mbar. MC(H35) → 1.25 K. Still(P1) → 10 mbar.
MC temperature wiggles as it passes through the Al Tc
1805 V6 → 13.8%. Condenser(G1) peaks at 255 mbar. MC(H35) → 0.84 K. Still(P1) → 3.0 mbar.
1815 V6 → 14.4%. Condenser(G1) peaks at 188 mbar. MC(H35) → 1.7 K. Still(P1) → 0.62 mbar.
1820 V6 → 15.0%. Condenser(G1) peaks at 178 mbar. MC(H35) → 0.72 K. Still(P1) → 0.20 mbar.
1835 V6 → 100%. Condenser(G1) peaks at 152 mbar. MC(H35) → 52 mK. Still(P1) → 0.15 mbar.
1850 Since the 1K Pot is running a bit low, spent an hour opening V14 in bursts to let the rest of the mix out of the dump and into circulation, keeping the Still pressure below 40 mbar. Repeated about 20 times before leaving it open. Little peaks in the condenser pressure and MC temperature.
1940 1K Pot rises from 1.56K/8.5mbar → 1.59K/12±1mbar, good.
1950 Still = 0.90 mbar.
Close V14. Still → 0.11 mbar. MC → 29.1 mK.
2100 MC = 28.3 mK. Still = 1.12 K / 0.11 mbar. G1/G2 = 120/108 mbar. 1K Pot = 1.59 K / 12±1 mbar.
2255 Check phonon channels. Get all but DCRC3pA back to -0.80 baselines. Adjust DCRC3pA offset to get its baseline close to -0.80V too.
MC rises to 29.3 mK after turning on DAQ programs.
2256 Dial HV up to -109.25V nominal, -107.2V actual, for 15% prebias (using RootSequencer)
2301 Dial HV down to -95V nominal, -92.9V actual
2308 Vdet is 89%. Adjust HV down to -94V nominal.
2310 HV = -91.8 V, Vtest = -21.2 mV, Ileak = -23.3 uA, Vdet = -82.5 V (90%).
Check triggering. Looks like the low DCRC1pA threshold of 2 still works fine, plenty of pulses showing up in the right place.
2314 Start -82.5 V PuBe data series ## min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pA threshold 2.
Start series with the Start Button
2335 HV = -91.8 V, Vtest = -21.1 mV, Ileak = -23.2 uA, Vdet = -82.6 V (90%).