2022/09/12 (Monday)

Previous Day

  • Lab in-personnel: Zach
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Warmed up yesterday, Sixth 5K pumpout.
1136 MC = 5.70K, MB = 37.5%
1137 Fill LN2 cold trap bath
LN2 dewar ran out, will need to be refilled
1139 Turbo on IVC = 1.4 nbar. With Turbo valved out for 1 minute IVC = 12 nbar.
1141 Close IVC valves on top of fridge
1145 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (speedivalve closed).Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.23 bar on mechanical gauge.
1150 Pump out 1K Pot in stages, alternating 1K Pot Speedivalve and V4a, opening V4a in brief spurts to keep 4He pump pressure low (below 5 mbar).
1200 Start pumping 1K Pot continuously
1200 MC = 5.68 K, cooling slowly. 1K Pot = 3.39 K / 2.0 mbar, cooling slowly
1210 Open 1K Pot needle valve → 15°. 1K Pot → 3.04 K / 6.5 mbar
1220 Open 1K Pot needle valve → 30°. 1K Pot → 2.79 K / 9.0 mbar
1225 Open 1K Pot needle valve → 45°. 1K Pot → 2.46 K / 13 mbar
1225 G1/G2 = 383/356 mbar
1230 Take LN2 dewar to be filled
1250 MC/1KPot = 2.87/2.00 and dropping linearly
1345 Finish filling LN2 cold trap dewar
1346 MC/1KPot = 1.96/1.53, G1/G2 = 47.8/47.2 mbar
1410 Begin LHe transfer @ MB = 33.6%
1419 Close V3 and V9, leaving 59 mbar mixture in dump
1419 Turn on 3He pump
1425 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 257 mbar. MC(H35) → 1.43 K. Still → 1.47 K / 32 mbar.
1434 V6 → 13.0%. Condenser(G1) peaks at 270 mbar. MC(H35) → 1.37 K. Still(P1) → 1.41 K / 16 mbar.
1445 V6 → 13.4%. Condenser(G1) peaks at 235 mbar. MC(H35) → 1.18 K. Still(P1) → 1.19 K / 4.5 mbar.
1454 Ended LHe transfer @ MB = 95.6%
1457 V6 → 13.8%. Condenser(G1) peaks at 193 mbar. MC(H35) → 0.87 K. Still(P1) → 1.12 K / 1.7 mbar.
1505 G1 still rising slowly, now at 197
1513 V6 → 14.4%. Condenser(G1) peaks at 202 mbar. MC(H35) → 0.49 K. Still(P1) → 1.13 K / 0.95 mbar.
1520 V6 → 15.0%. Condenser(G1) peaks at 197 mbar. MC(H35) → 0.30 K. Still(P1) → 1.13 K / 0.55 mbar.
1530 V6 → 100%. Condenser(G1) peaks at 185 mbar. MC(H35) → 0.11 K. Still(P1) → 1.09 K / 0.13 mbar.
1545 MC = 54.5 mK, plug in DCRCs 1 and 3
MC temp stalls for a minute then continues to drop
1632 MC = 28.2 mK and pretty flat, Condenser = 136 and pretty flat. 1K Pot = 1.57 K / 12±1 mbar.
1634 Open V14 to pull the rest of the mixture out of the dump into circulation. (Accidentally opened V4 to dump for a second, didn't seem to cause any problems though)
MC rises to 66 mK, Still peaks above 30 mbar then falls again, Condenser rises to 232 mbar. 1K Pot unaffected.
as usual we'll wait till Still is below 1 mbar then close V14.
1700 Still = 0.6 mbar. close V14
1910 Check phonon pulses - looking good, baselines all around -0.8 V
1920 MC = 28.3 mK
we'll try a higher bias again. Lately we haven't been able to go very high without significant current leakage
Let's try -100V nominal, see how the leakage looks
1924 Dial HV up to -115V nominal, -112.9V actual, for 15% prebias, using RootSequencer
Vtest fell as low as -16 mV but then after a little under 5 minutes suddenly rose to a new plateau, same as last few times I've tried
1932 Dial HV down to -100V nominal, -97.9V actual
1935 HV = -97.8 V, Vtest = -28.4 mV, Ileak = -31.3 uA, Vdet = -85.3 V (87%). Not quite as bad as last time (Aug 29)
1942 Dial HV down to -71V nominal, -68.8 V actual
1952 Vtest has settled
HV = -68.8 V, Vtest = -4.4 mV, Ileak = -4.8 uA, Vdet = -66.8 V (97%).
1955 Pulses still look good
1955 Start -66.8 V PuBe data series 70 min long, SiHV-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
2107 Start -66.8 V PuBe data series 657 min long, NaI-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
2110 HV = -68.8 V, Vtest = -4.3 mV, Ileak = -4.8 uA, Vdet = -66.8 V (97%).
2110 MC = 28.6 mK