start › cdms › k100 › run_diary › 20220909

2022/09/09 (Friday)

Previous Day

  • Lab in-personnel: Giulia
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Cooled to base yesterday after fifth 5K pumpout. Temperature spikes returned as usual, this time at 7:50 am
0925 Plug DCRCs 1&3.
0930 Fill LN2 cold trap dewar
0940 MC/Coldplate = 27.1/148 mK, Still = 1.12 K / 0.080 mbar / 0 mW, G1/G2 = 144/143 mbar, 1K Pot = 60° / 1.59 K / 21±1 mbar, MB = 75.8 %
1045 Restart MIDAS and DAQ programs (after lots of disabling/reenabling DCRCs 1,3,4)
1110 Pulse viewer (SQUIDtuningGUI) isn't working. Recently our main data directories /data/chocula and /data/cdms1 were migrated to new disks by IT. Computers were rebooted to remount the networked drives. Now when I try SQUIDtuningGUI on cdms1 I get an error, something about memory corruption.
1115 Start 0V test series, SiHV-triggered (DCRC1pB 10).
Checked traces using BatViewer. Pulses look OK. One channel has a shifted offset. Can look at the baseline in raw data pulses and convert to Volts with following conversion: baseline/6553.6-1.25
Adjust offsets
1200 Start 0V test series, SiHV-triggered (DCRC1pB 10).
looks like baselines are all near -0.8 V again (2949 bins)
1256 Start 0 V PuBe data series 45 min long, SiHV-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
1340 Start 0 V PuBe data series 182 min long, both NaI-&SiHV-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB/DCRC4pA threshold 10/20.
Start series with the Start Button
Had both triggers set in this series by mistake.
1648 Start 0 V PuBe data series 196 min long, NaI-triggered, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20.
Start series with the Start Button
1702 Between 5:02pm and 5:20pm MC temperature wandered up to 63 mK then back down. Should see if that has an effect on the data.
2010 Dial HV up to -27.6V nominal, -25.3V actual, for 15% prebias (using RootSequencer)
note: forgot to open LN2 Dewar vent for Ebox purge. We'll see how bad the leakage is…
actually checking later it looks like it was open and flowing.
2016 Dial HV down to -24.0V nominal, -21.7V actual
2023 HV = -21.7 V, Vtest = -2.2 mV, Ileak = -2.4 uA, Vdet = -20.7 V (96%). (good)
2023 Start -20.7 V PuBe data series 38 min long, SiHV-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
2025 Check pulses with pulse viewer, they look good
2102 Start -20.9 V PuBe data series 25 good min long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
2110 HV = -21.7 V, Vtest = -1.8 mV, Ileak = -2 uA, Vdet = -20.9 V (96%).
TowerFE crashed, need to restart series
2243 Start -20.9 V PuBe data series 796 min (13.3 hr) long, NaI-triggered with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC1pB threshold 10.
Start series with the Start Button
2245 HV = -21.7 V, Vtest = -1.7 mV, Ileak = -1.9 uA, Vdet = -20.9 V (97%).