UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20220905

2022/09/05 (Monday)

Previous Day

  • Lab in-personnel: Giulia
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4)
Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors.
PuBe behind lead wall on West side of lead/poly shield.
NaI trigger window: 7.6-11.2 MeV
Cooled to base yesterday after fourth 5K pumpout.
MC reached a minimum of 26.8 mK overnight. Spikes started at 6am.
1130 MC base is 27.1 mK with spikes peaking as high as 28.9
1130 Fill LN2 cold trap bath
1135 Plug DCRCs 1&3
1145 MC/Coldplate = 27.1/150 mK, Still = 1.12 K / 0.110 mbar / 0 mW, G1/G2 = 156/144 mbar, 1K Pot = 60° / 1.60 K / 15±1 mbar, MB = 62.6 %
1750 MC base is 28.4 mK with spikes peaking as high as 31.5

Next Day