2022/07/02 (Saturday)
- Lab in-personnel: Giulia
| Status | Run 76 in progress |
|---|---|
| Setup | Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRCX(4) Lead/poly shield fully assembled. Boron-filled socks covering NaI detectors. PuBe source behind lead wall. NaI trigger window -t10760 -T15760 (7.6-11.2 MeV) |
| 0905 | Check phonon channels. Relock DCRC3pA&B. DCRC3pB has SQUID locks back to where it was a couple weeks ago: offset 0.4 → baseline ~ -0.8. DCRC3pA required an offset change, 0.55→ 0.3. |
| 0911 | Start 0 V PuBe data series 65 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10. Start series with the Start Button. |
| 1020 | MC = 29.1 mK |
| 1030 | MC/Coldplate = 29.0/152 mK, Still = 1.08 K / 0.095 mbar / 0 mW, G1/G2 = 131/125 mbar, 1K Pot = 60° / 1.55 K / 8.0±0.5 mbar, MB = 74.1% |
| 1037 | Filled LN2 cold trap dewar |
| 1047 | Dial HV up to -227.1 V using RootSequencer for prebias. Set it to rise about 5 V/s with 0.25 second steps. |
| 1052 | After 5 minutes dial HV down to -197.3 V |
| 1055 | Start -197.3 V PuBe data series 171 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC4pA threshold 20. Start series with the Start Button. |
| 1105 | Check Vtest on the HV board: around -79 or -80 mV. Shoot, this is still too high - pulses look better today but leakage is still bad. |
| 1110 | Check again the purge gas flow at the Vent port on the LN2 Dewar. Didn't seem like there was any flow. Open valve more to restart flow |
| 1110 | Check Vtest again: -66 mV, better. |
| 1115 | Vtest = -44 mV, better |
| 1118 | Vtest = -5 mV, much better |
| 1130 | Vtest = -2 mV, better |
| 1345 | Something is wrong with the data. Dump sizes were around 7M till Dump 11, then rate fell off during Dump 12, then after that no traces, dumps only 394K. But randoms are on so there should always be something, DAQ problem presumably. Gonna stop the series. |
| 1347 | Try stopping the run, MIDAS is hung up trying to stop |
| 1400 | Getting an error trying to start a new run, DCRC_driver and Tower_fe crash. Gonna restart MIDAS |
| 1405 | Crap - restarting MIDAS didn't help, getting the same problem |
Messages 14:04:56.410 2022/07/02 [triggerfe2,INFO] Client 'towerfe3_01' on buffer 'SYSTEM' removed by cm_watchdog because process pid 2035980 does not exist 14:04:56.408 2022/07/02 [triggerfe2,INFO] Client 'dcrc_driver01' on buffer 'SYSTEM' removed by cm_watchdog because process pid 2035954 does not exist 14:04:54.806 2022/07/02 [mhttpd,INFO] Run #47581 start aborted 14:04:54.804 2022/07/02 [mhttpd,ERROR] [midas.c:4444:cm_transition,ERROR] Could not start a run: cm_transition() status 503, message '(null)' 14:04:54.703 2022/07/02 [mhttpd,ERROR] [midas.c:5730:cm_shutdown,ERROR] Killing and Deleting client 'dcrc_driver01' pid 2035954 14:04:54.702 2022/07/02 [mhttpd,ERROR] [midas.c:5728:cm_shutdown,ERROR] Client 'dcrc_driver01' not responding to shutdown command 14:04:54.701 2022/07/02 [mhttpd,ERROR] [midas.c:5730:cm_shutdown,ERROR] Killing and Deleting client 'towerfe3_01' pid 2035980 14:04:54.699 2022/07/02 [mhttpd,ERROR] [midas.c:5728:cm_shutdown,ERROR] Client 'towerfe3_01' not responding to shutdown command 14:04:49.792 2022/07/02 [mhttpd,ERROR] [midas.c:3660:cm_transition_call,ERROR] cannot connect to client "dcrc_driver01" on host cdms1.spa.umn.edu, port 25390, status 503 14:04:49.789 2022/07/02 [mhttpd,ERROR] [midas.c:9321:rpc_client_connect,ERROR] timeout on receive remote computer info: 14:04:49.787 2022/07/02 [mhttpd,ERROR] [midas.c:3660:cm_transition_call,ERROR] cannot connect to client "towerfe3_01" on host cdms1.spa.umn.edu, port 26873, status 503 14:04:49.785 2022/07/02 [mhttpd,ERROR] [midas.c:9321:rpc_client_connect,ERROR] timeout on receive remote computer info:
| 1420 | Tried disabling all but DCRC1 and restarting DAQ programs - same behavior. Tried disabling all but DCRC3 - was able to start a run OK that way. hmm. |
|---|---|
| 1425 | With DCRC3,2,4 all enabled, was able to start DAQ programs and start a series with no problem. |
| 1425 | Try again with all DCRCs enabled, same problem. So maybe we need to powercycle DCRC1. |
| 1435 | Dial HV down to 0V. This was a little tricky. At first I tried this with all DCRCs enabled including DCRC1, but I guess running the RootSequencer produced the same problem - HV didn't change and DAQ programs stopped. I looked at the ODB and it looked like DCRC2 qI was set to >11, so the commands went to the ODB but didn't go to the DCRC. Then I got the programs started again with DCRC1 disabled and I think that suddenly reset DCRC2qI, so the bias suddenly went to 0V. |