2022/05/02 (Monday)

Previous Day

  • Lab in-personnel: Matt
Status Run 76 in progress
Setup Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRC52(6)
1215 MC/Coldplate = 28.5/145 mK, Still = 1.09 K / 0.100 mbar / 0 mW, G1/G2 = 118/111 mbar, 1K Pot = 18° / 1.57 K / 10±1 mbar, MB = 32.6%
NOTE: we've been using Omni_Logger TestCopy_Himangshu.vi to get the Still and Coldplate temperatures. But since the beginning of the run this has given a Coldplate temperature of exactly 140 mK every time with no variation. Today I decided to look up the calibration curves and manually calculate the temperatures from the resistance readouts on the bridge. Still was the same as the vi but Coldplate was a little different, 146 mK. I found that the LakeShore bridge needs to have it's channel manually switched in order to get a fresh reading from the vi. I think the vi was just getting the last scanned value all this time…
1245 Fill LN2 cold trap bath
1400 Reset the HV maximum. First, unplug HV cable from HV box. Next, use RootSequencer to dial to current HV max (-250 V) by dialing DCRC2qI down from 11.64 to 0 V. Next, on Kepco supply dial HV up to -350 V - that's the new maximum. Note: that's too high to read on the HP supply (max 300 V), so instead I used the Fluke multimeter to set it to 350. Finally dial HV back down to 0V using the RootSequencer.
1410 The DCRC1qI value to get to HV = 0 has shifted a little, previously 11.64 now 11.66. (Not exactly zero, around HV = -45 mV.)
1507 Begin LHe transfer @ MB = 28.8%
1553 End LHe transfer @ MB = 95.6%