2022/04/28 (Thursday)
- Lab in-personnel: Matt, Nikki
| Status | Run 76 in progress |
|---|---|
| Setup | Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRC52(6) |
| 1440 | MC/Coldplate = 28.5/140 mK, Still = 1.09 K / 0.100 mbar / 0 mW, G1/G2 = 110/104 mbar, 1K Pot = 18° / 1.57 K / 10±1 mbar, MB = 36.9% |
| 1440 | Goldman = 72.2 mK @ 632 kOhm excitation |
| 1445 | Fill LN2 cold trap bath |
| 1545 | Begin LHe transfer @ MB = 35.3% |
| Tried weighing the 250L Dewar several times, to see how off we could potentially be. In the lightest case one wheel was maybe slightly touching the floor. In the heaviest case one bottom corner was touching the wall. The measurements spanned 1.7 pounds. Preferred procedure: make sure wheels aren't touching floor and corners aren't touching wall, readjust a few times (rotate a little against wall bar, push down, pull up), take heaviest measurement. | |
| 1627 | End LHe transfer @ MB = 95.6% |
| Gonna redo -180V data cuz it looks like an outlier | |
| 1651 | Dial HV up to -208.1 V using RootSequencer for prebias. Set it to rise about 5 V/s with 0.25 second steps. |
| 1658 | After 5 minutes dial HV down to -180.8 V |
| 1659 | Start -180.6 V data series after 5 min 13% prebias, 54 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pD threshold 10. Start series with the Start Button. |
| 1700 | Tower = 28.2 mK, a bit cooler than during other recent series |
| 1755 | Dial HV down to 0V after data series ends. |