2022/04/23 (Saturday)
- Lab in-personnel: Giulia
| Status | Run 76 in progress |
|---|---|
| Setup | Cables 2/3/1 → diverter/nothing/nothing → Side2-readout HV CONN 1/3/2 → HV 1(L)/2(R) → DCRC50(3)/DCRC55(1), DCRC33(2)→HV control, NaI trigger signal → DCRC52(6) |
| 1257 | Dial HV up to -91.8 V using RootSequencer for prebias. Set it to rise about 5 V/s with 0.25 second steps. |
| 1302 | After 5 minutes dial HV down to -79.6 V |
| 1305 | Hmm, baselines seem significantly elevated and pulses in some channels look weak. Doesn't look as good as the recent 100V and -100V data. Maybe the warm humid day is causing leakage at the e-box and we need to start purging it. |
| 1310 | Dial HV down to 0V. |
| 1315 | Baselines at -80V: -0.765, -0.686, -0.763, -0.695, -0.660, -0.654 Baselines at 0V: -0.900, -0.826, -0.876, -0.877, -0.843, -0.810 |
| 1535 | MC/Coldplate = 28.3/140 mK, Still = 1.10 K / 0.110 mbar / 0 mW, G1/G2= 108.7/122 mbar, 1K Pot = 18° / 1.57 K / 18±1 mbar, MB = 68.3% |
| 1535 | Goldman = 73.2 mK @ 632 uV excitation |
| 1540 | Fill LN2 cold trap bath |