start › cdms › k100 › run_diary › 20220221

2022/02/21 (Monday)

Previous Day

  • Lab in-personnel: Matt, Michael, Shubham
Status Run 75 finished. Warmup for last several days. 1K Pot pumped dry and isolated. All mixture in dump, fridge isolated. LHe Trap and LN2 Trap 1 still have contaminants. ~1 atm exchange gas in IVC. Cryostat Vacuum Jacket pumping out overnight. Main bath empty and disconnected from recovery.
1310 MC(Cernox) = 295 K
1315 Turbo on Cryostat Vacuum Jacket: 25 ubar. With Turbo valved out for 1 minute, CVJ = 250 ubar.
1340 Stop pumping CVJ, close valve and remove hose.
Remove all three sets of main bath bolts, then lower MB into pit.
1400 Erect clean tent at bottom of fridge.

analysis note:

  • tried the same calibration and noise width determination technique on Run68 data that I've used for Run75.
    • first observation: saturation is much weaker with the -125V data with S101 than with S104. Linearizing OF wrt INT seems hardly necessary
    • noise width = 8.3 eVee. So we definitely seem to be doing better with S104.