2022/02/15 (Tuesday)

Previous Day

  • Lab in-personnel: Giulia
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board+extra_board/nothing → Side2-readout HVboard CONN1/2/3 → HV1/2 → DCRC55(1)/50(3)
Lead shield in place. Am Source y = 0.03 mm
Testing Side2-readout HV board
0850 Goldman = 41.5 mK
0910 Mini-IbIs on DCRC1pA; adjust QET bias for dP/dI = 0.4 uA
0935 Mini-IbIs on DCRC1pB; adjust QET bias for dP/dI = 0.4 uA
0945 Mini-IbIs on DCRC1pD; adjust QET bias for dP/dI = 0.4 uA
0955 Mini-IbIs on DCRC3pA; adjust QET bias for dP/dI = 0.4 uA
1005 Mini-IbIs on DCRC3pB; adjust QET bias for dP/dI = 0.4 uA
1020 Mini-IbIs on DCRC3pD; adjust QET bias for dP/dI = 0.4 uA
1020 Goldman = 41.3 mK
1022 Dial HV up to 113.2 V for prebias
1025 Adjust driver offsets to set baselines around -0.9 V
1028 After 5 minutes of prebias dial HV down to 98.6 V
1030 Start 98.62 V data series after 5 min 15% prebias, 57 min long, with 1000 initial randoms, 5 Hz in-run randoms, trigger on DCRC3pB threshold 10.
Start series with the Start Button.
Bias pretty stable, 98.6-98.63 V
Noticed a couple of times now that the Goldman temperature becomes a little noisier when the bias is on.
1130 Dial HV down to 0V after series ends
1315 Goldman/Tower/MC(H35)/Coldplate = 36.8/97.1/32.4/175 mK, Still = 1.30 K / 0.130 mbar / 0 mW, G1/G2= 290/267 mbar, 1K Pot = 1.58 K / 11±1 mbar, MB = 30.9%
1327 Fill LN2 cold trap bath.