UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20220124

2022/01/24 (Monday)

Previous Day

  • Lab in-personnel: Giulia and Matt
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4)
Lead shield in place. Am Source y = 20.01 mm
1034 Goldman/Tower/MC(H35)/Coldplate = 42.0/80.5/34.1/180 mK, Still = 1.43 K / 0.110 mbar / 0 mW, G1/G2= 265/250 mbar, 1K Pot = 1.60 K / 14±1 mbar, MB = 31.4%.
1048 Fill LN2 cold trap bath
Sensitivity of standard QET bias setting to temperature
Channel uA/mK
DCRC1pA -0.76
DCRC1pB -
DCRC1pC -1.21
DCRC1pD -1.97
DCRC3pA -1.96
DCRC3pB -1.13
DCRC3pC -0.36
DCRC3pD n/a
DCRC4pA -2.69
DCRC4pB -1.73
DCRC4pC -1.12
DCRC4pD -1.27
1330 Goldman = 41.4 mK. Set QET biases to target dP/dI=1.0 uV and adjust driver offsets so baselines ~ -0.9 V
1339 Start 0V data series, 40 min long, with 1000 initial randoms, no in-run randoms, trigger on DCRC3pB threshold 10.
Start series with the Start Button.
1610 Begin LHe transfer @ MB = 22.4%
1620 Double-click Arduino Bridge link on desktop. Turn off GwINSTEK power output, then establish communication with the Arduino. (No temperature spike because I turned off the power output.) The turn GwINSTEK power output back on.
1625 Set Dwell to 40 ms
1632 Do mvNwait +4steps×151 times, 40 second wait, with dwell = 40 and power = 6.8 V. (Around 5.93 mm/hour)
This sequence will take 1hr40min00sec and will move the source a total distance of 9.96 mm ending at y = 10.05 mm. Should end at 6:12 pm.
Tower temperature averages around 0.51 K with familiar oscillations.
1700 End LHe transfer @ MB = 93.1%
2010 Goldman = 45 mK and cooling
Let's measure Tc's on DCRC3 channels.
2020 Worked out again how to edit the ODB from the command line, by entering commands like odbedit -c 'set “/Equipment/Tower01/Settings/DCRC4/Phonon/QETBias (uA)[1]” 0.0'
2030 Set all QET biases to 0
2040 Set up DCRC3 for Tc tests. Set Test Signal to 12 mVpp / 200 Hz. Trigger on test signal. Connect test signal to QETs.
Modulation is 10.7/14.5/15.0 mVpp in pA/pB/pC Goldman = 44.0 mK.
2042 Turn on DCRC1 LEDs with rep rate reduced from 5000 to 500. Turn off LEDs after 3 minutes. Goldman peaked at 1.28 K.
2045 Modulation is 3.0/3.5/2.8 mVpp in pA/pB/pC Goldman = 350 mK.
2110 See transitions in DCRC3pA/pB/pC @ 70/69/77 mK
2120 Set QET biases back into transition and relock SQUIDs where necessary
2125 Goldman = 47.7 mK and cooling