2022/01/10 (Monday)

Previous Day

  • Lab in-personnel: Matt
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4)
Lead shield in place. Am Source y = 20.01 mm
1020 Goldman = 41.2 mK. Drifting down slowly overnight from 41.4 mK.
Let's do another set of automated IbIs at this much lower temperature
1035 Disable DCRC3&4 on ODB. We're only taking data with one DCRC for IbIs.
1040 Adjust DCRC1pA driver offset so that IbIs peak (QET bias around 290) is near top of output scale (around 1V).
oops, had the logger on, so some junk data was saved
1055 Start automated IbIs on DCRC1pA. 550 to 250 uA step -2.
1055 Goldman = 41.2 mK
1130 Make new directory in /byseries and move the QB_scan files from /livedata to the new directory.
1135 Adjust DCRC1pB driver offset so that IbIs peak (QET bias around 320) is near top of output scale (around 1V).
1139 Start automated IbIs on DCRC1pB. 500 to 200 uA step -2.
1140 Goldman = 41.2 mK
1210 Make new directory in /byseries and move the QB_scan files from /livedata to the new directory.
1220 Adjust DCRC1pC driver offset so that IbIs peak (QET bias around 330) is near top of output scale (around 1V).
1220 Start automated IbIs on DCRC1pC. 500 to 200 uA step -2.
1220 Goldman = 41.3 mK
1300 Make new directory in /byseries and move the QB_scan files from /livedata to the new directory.
1305 Adjust DCRC1pD driver offset so that IbIs peak (QET bias around 210) is near top of output scale (around 1V).
1308 Start automated IbIs on DCRC1pD. 400 to 100 uA step -2.
1310 Goldman = 41.5 mK
1500 Make new directory in /byseries and move the QB_scan files from /livedata to the new directory.
The range chosen for DCRC1pA auto IbIs was too high, need to redo it
1525 Adjust DCRC1pA driver offset so that IbIs peak (QET bias around 290) is near top of output scale (around 1V).
1531 Start automated IbIs on DCRC1pA. 500 to 200 uA step -2.
1535 Goldman = 41.5 mK
1715 Make new directory in /byseries and move the QB_scan files from /livedata to the new directory.
1700 Goldman/Tower/MC(H35)/Coldplate = 41.7/77.8/33.3/172 mK, Still = 1.30 K / 0.110 mbar / 0 mW, G1/G2= 248/211 mbar, 1K Pot = 1.61 K / 17±1 mbar, MB = 25.6%.
1710 Begin LHe transfer @ MB = 25.4%
1715 Fill LN2 cold trap bath
1741 End LHe transfer @ MB = 91.8%
2130 Restart DAQ programs with only DCRC3 enabled
2135 Adjust DCRC3pA driver offset so that IbIs peak (QET bias around 190) is near top of output scale (around 1V).
2142 Start automated IbIs on DCRC3pA. 400 to 100 uA step -2.
2145 Goldman = 41.9 mK
2215 Make new directory in /byseries and move the QB_scan files from /livedata to the new directory.
2220 Adjust DCRC3pB driver offset so that IbIs peak (QET bias around 180) is near top of output scale (around 1V).
2225 Start automated IbIs on DCRC3pB. 400 to 100 uA step -2.
2225 Goldman = 41.8 mK
2300 Make new directory in /byseries and move the QB_scan files from /livedata to the new directory.