start › cdms › k100 › run_diary › 20220109

2022/01/09 (Sunday)

Previous Day

  • Lab in-personnel: Michael
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4)
Lead shield in place. Am Source y = 20.01 mm
Yesterday I turned 15V power off on DCRCs3&4, left on on DCRC1 - this seems to have made the tower significantly cooler.
0945 Michael arrive in lab.
0950 LN2 Fill
0950 Goldman/Tower/MC(H35)/Coldplate = 41.7/76/33.8/179 mK, Still = 1.433±0.002 K / 0.110 mbar / 0 mW, G1/G2= 237/204 mbar, 1K Pot = 1.61 K / 18±1 mbar, MB = 70.8%.
1007 Michael leaves lab.
With temperature now lower I did mini-IbIs on all channels. Had to zap some channels because some weren't keeping lock and sometimes the temperature was drifting up. Getting the channels more stable got the temperature down.
Gonna take a second series at this source position to see how different it looks now that the temperature is lower
2130 Goldman = 41.6 mK. Set QET biases targeting dP/dI = 1.0 uV
2139 Start 0V data series, 40 min long, with 1000 initial randoms, no in-run randoms, trigger on DCRC3pB threshold 10.
Start series with the Start Button.
2144 Goldman = 41.5 mK. Reset QET biases targeting dP/dI = 1.0 uV