| Status | Run 75 in progress |
| Setup | Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4) |
| 1115 | Goldman/Tower/MC/Coldplate = 50.0/91.3/36.7/219 mK, Still = 1.67±0.04 K / 0.130 mbar / 0 mW, G1/G2= 184/181 mbar, 1K Pot = 1.64 K / 21±2 mbar. MB = 67.6% |
| | Coldplate and Still thermometers are still reading high, still thermometer reading not very stable. |
| | Goldman thermometer is now cooler than its calibrated range. The temperature above is based on an extrapolation of the coldest 5 points in the calibration curve. I fit 1/T versus R to a quadratic equation to fit the data. |
| 1120 | Fill LN2 cold trap bath. |
| 1150 | Modify Omni_LoggerCopy.vi to include the extrapolation of the Goldman thermometer |