2021/12/15 (Wednesday)

Previous Day

  • Lab in-personnel: Matt
Status Run 75 in progress
Setup Tower Faces 5/3/2 → Cables 2/1/3 → nothing/diverter_board/nothing → DCRC55(1)/DCRC50(3)/DCRCx(4)
1050 Goldman/Tower/MC/Coldplate = 50.3/111/36.7/221 mK, Still = 1.70±0.04 K / 0.130 mbar / 0 mW, G1/G2= 175/182 mbar, 1K Pot = 1.64 K / 26±2 mbar. MB = 33.4%
Coldplate and Still thermometers are still reading high, still thermometer reading not very stable.
1100 Fill LN2 cold trap bath.
1145 Got yesterday's DCRC4pA IbIs to process (the code had a dumb bug that only allowed “detector” numbers up to 3).
Still need to move yesterday's DCRC4pB IbIs files to new directory
1155 Make new directory in /byseries and move the QB_scan files from /livedata to the new directory.
1200 Adjust DCRC4pC driver offset so that IbIs peak (QET bias around 340) is near top of output scale (around 1V).
1201 Start automated IbIs on DCRC4pC. 500 to 200 uA step -2.
1245 Make new directory in /byseries and move the QB_scan files from /livedata to the new directory.
1245 Noticed that temperature (Goldman thermometer) increased from 50.2 to 50.5 mK starting at 12:30. Took a look at DCRC4 channels - pC was railed, relocked it, looked like it may be in s/c, upped the QET bias from 300 to 350 to see pulses again. After that temperature started falling again.
1300 Adjust DCRC4pD driver offset so that IbIs peak (QET bias around 210) is near top of output scale (around 1V).
1306 Start automated IbIs on DCRC4pD. 400 to 100 uA step -2. MIDAS got stuck, this only got down to 146 uA (see below)
1340 Looks like MIDAS got stuck. Browser interface isn't responding. Data files were showing up a few per minute until 13:33, then they stopped.
1345 Kill then restart MIDAS. Restart DAQ programs. Sequencer appears to be stuck - stuck on Transition Start line, “Cannot start run” error message. Not sure how to get out, the Stop Immediately button isn't there…
Clicking “Load sequence” brings up the ordinary page. I guess the sequence terminated itself. Need to start over I guess. \\… although sequence got down to 144 uA, that may be enough.
1350 MIDAS is in some odd state - can't start data… wait I fixed this by turning off the logger.
may need to re-run the sequence just to get the proper stuff reset.
1400 Make new directory in /byseries and move the QB_scan files from /livedata to the new directory.
1505 Run a short IbIs sequence just to let the sequencer perform all the reset steps at the end. Delete the data files.
1515 Begin LHe transfer @ MB = 26.4%
1551 End LHe transfer @ MB = 90.4%

Next Day