| 1115 | Perform mixture throughput test at MC = 77 K using above procedure with 3He pump on. Let G1 rise to 543 mbar before Step 3, then it falls to 69 mbar, then rises. Throughput = 42.4 mbar/min. Nearly the same as last time, 2021 June 10 |
| 1120 | Open mixture to both sides of the fridge and cold LN2 Trap 1. |
| 1120 | Turbo on IVC = 6.3 nbar. With Turbo valved out for 1 minute, IVC = 100 nbar. |
| 1125 | Try Leak Detector on IVC, with Turbo still pumping in parallel. (We've been pumping out the helium for 2 hr 40 minutes.) LD = 510 pbarL/s. With Turbo valved out for a minute, LD → 2200 pbarL/s. Not as good as last time. Open Turbo valve to continue pumping IVC. |
| 1150 | IVC = 470 pbarL/s with turbo, 2000 pbarL/s without turbo |
| 1155 | Perform electrical checks on all 3 cables. All looks good except: Cable 3 Feedback A too high, 0.8 MOhm, should be around 10 kOhm. |
| | This same feedback resistor read high at 77K the last time this SQUET, ZC-70, was used, in Run 71 - with iZIP G106a. Side1 pB in that run - it seems to have worked fine. |
| 1245 | IVC = 411 pbarL/s with turbo, 2000 pbarL/s without turbo |
| | The LHe liquefier was down the last couple weeks, but is now running again. We thought there'd be LHe today but we won't have any till tomorrow, so we have to delay one day. To keep the fridge at 77K we probably want to add more LN2 - not sure how long till it boils off. |
| 1510 | Tower/Sorb = 77.2/130.9 K |
| 1510 | Set up for another LN2 transfer. Insert level resistor and let the value stabilize. |
| 1510 | Level resistor = 1.235 kOhm. |
| 1515 | Start LN2 transfer. Level resistor immediately rises, → 1.305 kOhm. |
| 1520 | Level resistor = 1.335 kOhm. |
| 1525 | Fill LN2 cold trap bath. (Used the 5L Dewar, filled it in the fill room.) |
| 1535 | Sorb = 130.3 K (down a little), Level resistor = 1.332 kOhm. |
| 1550 | Looks like we ran out of LN2 in the Dewar. |
| 1550 | Sorb = 130.0 K, Level resistor = 1.286 kOhm. |
| 1600 | Close LN2 Dewar valve. Pull out Level Resistor and put stopper in port. Put ping-pong ball on MB exhaust port. |