2021/11/29 (Monday)
- Lab in-personnel: Matt
| Status | Beginning Run 75. |
|---|---|
| 1K Pot isolated. MB raised. Cryostat vacuum jacket (CVJ) pumped out. | |
| 1005 | Check contents of warm LN2 Trap 1: 6.4 mbar. Pump out with Scroll Pump. While pumping pressure falls by 5.7 mbar. |
| 1005 | Check contents of warm LN2 Trap 2 (which was pumped out 5 days ago): 2.3 mbar. Pump out with Scroll Pump. While pumping pressure falls by 0.9 mbar. interestingly Trap 2 doesn't seem to pump down as low as Trap 1, as seen on both G1 and P2. |
| 1040 | Start cryocleaning mixture through LN2 Cold Trap 2, V6 = 12.4% |
| 1045 | Increase V6 → 12.6% to restart flow |
| 1105 | Increase V6 → 12.7% to restart flow |
| 1115 | Increase V6 → 12.8% to speed up flow |
| 1130 | Increase V6 → 12.9% to speed up flow |
| 1140 | Stop cleaning through Trap 2 after 60 min. G2 flow pressure was 575-825 mbar. Pump mixture to dump. Removed 6.5(5.4) mbar. |
| 1205 | Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.8% |
| 1220 | Increase V6 → 12.9% to restart flow |
| 1245 | Increase V6 → 13.0% to speed up flow |
| 1315 | Stop cleaning through Trap 1 after 70 min. G2 flow pressure was 590-730 mbar. Pump mixture to dump. Removed 8.5(7.8) mbar. |
| 1420 | Lower MB into pit far enough to install fill funnel on fridge. Raise MB again. |
| 1605 | Check contents of warm LN2 Trap 2: 6.5 mbar. Pump out with Scroll Pump. While pumping pressure falls by 5.4 mbar. |
| 1615 | Check contents of warm LN2 Trap 1: 8.5 mbar. Pump out with Scroll Pump. While pumping pressure falls by 7.8 mbar. |
| 1615 | Turn off 3He pump. |