2021/09/19 (Sunday)

Previous Day

  • Lab in-personnel: Giulia and Matt
Status Run 74 in progress. First cooldown after IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. Lead shield erected, 4 walls. Currently both DCRCs are on.
Position of Am source: y = 0.0 mm
1323 Tower/MC/Coldplate = 113.5/38.8/199 mK, Still = 1.38 K / 0.095 mbar / 0 mW, G1/G2= 342.8/319.4 mbar, 1K Pot = 1.51 K / 10±1 mbar. MB = 32.6%.
1435 Fill LN2 cold trap bath
1342 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 10/3.0/2.4/46 8e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 81/12/27/330 e-6.
Noise is slightly higher than yesterday especially @3kHz, while some are slightly lower @10kHz.
1323 Start LHe transfer @ MB = 32.6%
1417 End LHe transfer @ MB = 91.3%
1444