2021/09/18 (Saturday)
- Lab in-personnel: Giulia
| Status | Run 74 in progress. First cooldown after IVC rebuild. |
|---|---|
| Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels. | |
| Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/ |
|
| Position of Am source: y = 0.0 mm | |
| 1047 | Tower/MC/Coldplate = 94/41.5/193 mK, Still = 1.39 K / 0.1000 mbar / 0 mW, G1/G2= 340.6/304.4 mbar, 1K Pot = 1.56 K / 10±1 mbar. MB = 69.6%. |
| 1059 | Fill LN2 cold trap bath |
| 1018 | At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 10/3.0/2.7/60 8e-6. This is q3/q1/q4/q2 At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 67/14/7.5/250 e-6. |
| Noise is slightly higher than yesterday especially @3kHz, while some are slightly lower @10khz. | |
| 1123 | Start 45 min series of +10/0V data (bias applied after 60 sec), at elevated temperature, with 1000 beginning randoms, trigger threshold 80/55/35/200 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Use Sequencer script r74_takeandflash_dcrcbias.msl. This flashes for 30 seconds at end of series. |
