2021/09/12 (Sunday)
- Lab in-personnel: Giulia
| Status | Run 74 in progress. First cooldown after IVC rebuild. |
|---|---|
| Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels. | |
| Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/ |
|
| Position of Am source: y = 0.0 mm | |
| 1403 | Tower/MC/Coldplate = 85.6/34.2/188.1 mK, Still = 1.27 K / 0.070 mbar / 0 mW, G1/G2= 291.15/263.8 mbar, 1K Pot = 1.56 K / 10±1 mbar. MB = 62.3%. |
| 1409 | Fill LN2 cold trap bath |
| 1411 | At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 8.8/2.5/2.3/18 8e-6. This is q3/q1/q4/q2 At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 310/19/34/520 e-6. |
| Noise is lower than yesterday. | |
| 1422 | Start 45 min series of +10V/0V, +12V/0V (bias applied after 60 sec and after 4min increased it to +12V), with 1000 beginning randoms, trigger threshold 100/60/30/160 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Use Sequencer script r74_takeandflash_dcrcbias.msl. This flashes for 30 seconds at end of series. |
| This data seems to have overtriggered. |