UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20210911

2021/09/11 (Saturday)

Previous Day

  • Lab in-personnel: Giulia and Michael
Status Run 74 in progress. First cooldown after IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. Lead shield erected, 4 walls. Currently both DCRCs are on.
Position of Am source: y = 0.0 mm
1336 Tower/MC/Coldplate = 92.1/35.2/189.9 mK, Still = 1.34 K / 0.070 mbar / 0 mW, G1/G2= 287.5/268.5 mbar, 1K Pot = 1.56 K / 10±1 mbar. MB = 31.9%.
1344 Fill LN2 cold trap bath
1008 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 16/4.4/3.7/90 8e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 900/32/85/1600 e-6.
Noise is higher than yesterday.
1405 Start LHe transfer @ MB = 31.9%
1443 End LHe transfer @ MB = 88.9 %
1515 Start 45 min series of +12V/0V (bias applied after 60 sec), with 1000 beginning randoms, trigger threshold 150/120/40/300 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Use Sequencer script r74_takeandflash_dcrcbias.msl. This flashes for 30 seconds at end of series.
This data seems to have overtriggered