UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20210831

2021/08/31 (Tuesday)

Previous Day

  • Lab in-personnel: Matt and Giulia
Status Run 74 in progress. First cooldown after IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. Lead shield erected, 4 walls. Currently both DCRCs are on.
Position of Am source: y = 0.0 mm
0843 Tower/MC/Coldplate = 143.9/35.6/191.5 mK, Still = 1.39 K / 0.070 mbar / 0 mW, G1/G2= 285/264 mbar, 1K Pot = 1.57 K / 10±1 mbar. MB = 73.6%.
0853 Fill LN2 cold trap bath.
0856 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 19/4.0/2.8/75 8e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 470/46/50/81 e-6
Noise is considerably higher than yesterday.
0941 Start 45 min series of +10V/0V, +9V/0V, +8V/0V, +7V/0V (initial bias applied after 60 sec, then 90 sec pauses to 9V, 8V, 7V), with 1000 beginning randoms, trigger threshold 120/40/40/400 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Use Sequencer script r74_takeandflash_dcrcbias.msl. This flashes for 30 seconds at end of series.
Reset bias using the Tools interface - this isn't very fast, took several seconds to change all 4 channels.
07210831_0941
Let's try heating the fridge again as on Aug 19. That time we wound up increasing the impact ionization strength at elevated temperature, then cooled again and saw the same increase (i.e. the heating seemed to induce a state change).
1225 Tower/MC = 162.5/ 35.1 mK
1225 Turn MC heater on to 600 μW
1320 Tower/MC = 0.871/0.884 K and seem to have settled. G1/G2 = 381/352 mbar, Still = 0.130 mbar.
1335 Start 45 min series of +10V/0V (bias applied after 60 sec), at elevated temperature, with 1000 beginning randoms, trigger threshold 150/70/40/400 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Use Sequencer script r74_takeandflash_dcrcbias.msl. This flashes for 30 seconds at end of series.
1445 Turn off MC heater
THIS LAST DATA SERIES DIDN'T WRITE! dammit MIDAS!
well… we can do another 10V series after it cools to base, and see if the stat has changed.
1545 Tower = 180 mK, seems to be pretty flat now. MC = 36.1 mK
1553 Start 45 min series of +10V/0V (bias applied after 60 sec), with 1000 beginning randoms, trigger threshold 120/50/40/450 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Use Sequencer script r74_takeandflash_dcrcbias.msl. This flashes for 30 seconds at end of series.

Next Day