2021/08/05 (Thursday)
- Lab in-personnel: Giulia
| Status | Run 74 in progress. First cooldown after IVC rebuild. |
|---|---|
| Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels. | |
| Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on. | |
| 1020 | Tower/MC/Coldplate = 140.05/35.3/191 mK, Still = 1.39 K / 0.075 mbar / 0 mW, G1/G2= 281.9/259.8 mbar, 1K Pot = 1.60 K / 20±1 mbar. MB = 69.3 %. |
| 1034 | Fill LN2 cold trap bath. |
| 1044 | At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 43/4.7/4.3/131e-6. This is q3/q1/q4/q2 At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 834/64/78/1430 e-6 |
| 1058 | Start 61 min series of 0V/-9V (bias applied after 60 sec), with 1000 beginning randoms, trigger threshold 120/40/35/500 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Start sequence with Start button. |
| 1200 | Dial bias down to 0V and flash detector for 10 minutes |
| Let's try moving the source. We'll move it to -10.0 mm. | |
| 1519 | Tried doing a move but the motor wires weren't plugged into the driver. |
| 1530 | Do mvNwait -4steps×152 times, 40 second wait, with dwell = 40 and power = 6.8 V. (Around 5.93 mm/hour) This sequence will take 1hour41min20sec and will move the source a total distance of 10.02 mm. Should end at 5:11 pm. Tower → 0.88/1.21/1.40 K after move 1/2/3. After move 12 Tower ~ 1.75 K, seems to be plateauing. After move 26, MC ~ 180 mK, Tower ~ 1.78 K |
| 1713 | Move sequence ends, slightly delayed. |
| 1725 | Tower/MC = 430/72 mK, cooling to base. |
| 1817 | MC = 36.2 mK, which is 1 mK above the eventual base temperature. 64 minutes after the end of the move sequence. |