2021/08/04 (Wednesday)

Previous Day

  • Lab in-personnel: Giulia
Status Run 74 in progress. First cooldown aftr IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on.
0930 Tower/MC/Coldplate = 212.5/36.6/197.3 mK, Still = 1.41 K / 0.075 mbar / 0 mW, G1/G2= 274.1/247.2 mbar, 1K Pot = 1.61 K / 20±1 mbar. MB = 34.3 %.
0948 Fill LN2 cold trap bath.
1316 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 30/4.2/3.6/85e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 740/111/73/2060 e-6
1419 Start 45 min series of 0V/-12V (bias applied after 60 sec), with 1000 beginning randoms, trigger threshold 150/50/40/700 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
1504 Dial bias down to 0V and flash detector for 10 minutes
1521 Begin LHe transfer @ MB = 25.6%
1601 End LHe transfer @ MB = 90.2%
07210804_1419

Next Day