start › cdms › k100 › run_diary › 20210720

2021/07/20 (Tuesday)

Previous Day

  • Lab in-personnel: Pramod and Giulia
Status Run 74 in progress. First cooldown aftr IVC rebuild.
Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on.
1101 Tower/MC/Coldplate = 160/35.6/194 mK, Still = 1.41 K / 0.075 mbar / 0 mW, G1/G2= 295/266.2 mbar, 1K Pot = 1.62 K / 20±1 mbar. MB = 66%.
1120 Fill LN2 cold trap bath.
1253 At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 32/2.95/2.72/107 e-6. This is q3/q1/q4/q2
At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 6191/91/39/6237 e-6
These are higher but some close to yesterday.
1305 Place Na-22 source 30 cm East of detector (center-to-center distance) on stand in lead castle.
1305 Increase ChargeWaveFormLength in DCRC1&3 from 8192→65536. This will increase the trace length from 1.6384 ms to 13.1072 ms.
1310 Apply -15V bias to Side 2. Now we are seeing some pulses.
1320 Event rate in DCRC1qI(q4) for pulses above 0.1V: 152 pulses counted in 200 traces. 58.0 Hz.
1325 Event rate in DCRC3qO(q1) for pulses above 0.1V: 107 pulses counted in 200 traces. 40.8 Hz.
1325 Dial bias down to 0V and flash detector for 60 seconds
1325 Restore ChargeWaveFormLength to standard settings.
1411 Start 50 min series of 0V/-15V (bias applied after 100 sec), with Na-22 source at 0.30 m, with 1000 beginning randoms, trigger threshold 100/50/40/180 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
1503 Dial bias down to 0V and flash detector for 60 seconds.
1555 Put Na-22 source back in safe
We'll try moving the source during a series again.
1604 Start 49 min series of 0V/-15V (bias applied after 60 sec), with 1000 beginning randoms, trigger threshold 100/50/40/180 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start sequence with Start button.
1619 Do mv -64, with dwell = 40 and power = 6.8 V. (Around 760 mm/hour)
This single move will take about 5 sec and will move the source a total distance of -1.06 mm.
Tower jumps from 92 mK to 3.43 K, then starts cooling. Traces look crazy for several seconds but then pulses come back - they maybe look smaller? This is of course a very different temperature.
1625 Tower = 1.0 K and falling
1629 Tower = 0.6 K and falling
1634 Tower = 0.4 K and falling
1640 Tower = 0.3 K and falling
1647 Tower = 0.25 K and falling
1702 Dial bias down to 0V and flash detector for 60 seconds

huh???

20210720_log.xlsx