2021/07/19 (Monday)
- Lab in-personnel: Pramod
| Status | Run 74 in progress. First cooldown aftr IVC rebuild. |
|---|---|
| Tower thermometer is reading high, probably this is a thermometer problem and the actual temperature is much lower. High noise in some of the charge channels. | |
| Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/Source_mover_board+HV_diverter_board → DCRC50(3)/55(1). Kepco 40V supplye → HV BNC. No lead/poly shield. Currently both DCRCs are on. | |
| 1101 | Tower/MC/Coldplate = 154/35.8/193 mK, Still = 1.40 K / 0.080 mbar / 0 mW, G1/G2= 300/268 mbar, 1K Pot = 1.62 K / 20±1 mbar. MB = 21.7%. |
| 1110 | Fill LN2 cold trap bath. |
| 1422 | Begin LHe transfer. MB = 17.1% |
| 1457 | Finish LHe transfer. MB = 90.3% |
| 1555 | At 10kHz noise in DCRC3qI/qO/DCRC1qI/qO = 38/6.0/4.2/110 e-6. This is q3/q1/q4/q2 At 3kHz (peak) noise in DCRC3qI/qO/DCRC1qI/qO = 6200/400/140/4300 e-6 |
| This is a lot worse than last Friday, the 3kHz oscillations are bad today. It's also very hot outside so likely more humid in the lab. | |
| 1610 | Place Na-22 source 40 cm East of detector (center-to-center distance) on stand in lead castle. |
| 1621 | Start series then abort - randoms set incorrectly. |
| 1623 | Start 45 min series of 0V/-15V (bias applied after 60 sec), with Na-22 source at 0.40 m, with 1000 beginning randoms, trigger threshold 180/70/40/450 on DCRC3qI/qO/DCRC1qI/qO, 5Hz in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Start sequence with Start button. |
| 1715 | Dial bias down to 0V and flash detector for 60 seconds. |